IMPORTANT!
Never change the structure of the bed or install on it
components other than those indicated in this document and
approved for the specific product by the manufacturer.
IMPORTANT!
Electrical and mechanical components (excluding the
accessories referred to in chapter 7 of the operating
instructions 'Installation and use of accessories') may only be
installed by service personnel authorized by Lojer service
(vocational upper secondary qualification in machine and
metal work or equivalent and valid SFS 6002 training in
occupational safety in electric work). Instructions for replacing
components by service personnel are given in these
maintenance instructions.
IMPORTANT!
When using a passenger hoist to lift the patient from the bed,
note the ground clearance and dimensions of the bed and
follow the instructions of the passenger hoist manufacturer.
IMPORTANT!
Do not attach mains cables of other devices or other wires to
the bed. When moving the bed or using its functions, make
sure that the cables of other devices are not left under the bed
wheels or between the bed's structures.
IMPORTANT!
Always disconnect the equipotential bonding before moving
the bed.
IMPORTANT!
The bed meets the electromagnetic compatibility requirements
of operating environment referred to in EMC directive
2004/108/EC. Make sure that the bed is not exposed to
electromagnetic radiation that exceeds the approved standard
limits. (EMC)
4.1 Safe use of power cable
Follow these safety instructions in the use of the power cable.
IMPORTANT!
Never tie the power cable to the bed, because a lifting
movement may cut it. Make sure that you can quickly unplug
the power cable in an emergency.
IMPORTANT!
Always unplug the power cable before moving the bed. Make
sure that the cable is not crushed between the bed's
structures or wheels.
IMPORTANT!
If the power cable is cut or damaged, unplug it immediately.
Only use the original power cable of the bed.