background image

TRID30-7

Temperature Recorder with Display 

and 30 Day Statistics Memory

Product User Guide

Document Release Version

:

 1.3

Published 1 September 2016

Copyright 

©

 LogTag

®

 Recorders  , 2004-2016

LogTag Recorders Ltd

Содержание TRID30-7

Страница 1: ...ID30 7 Temperature Recorder with Display and 30 Day Statistics Memory Product User Guide Document Release Version 1 3 Published 1 September 2016 Copyright LogTag Recorders 2004 2016 LogTag Recorders L...

Страница 2: ...10 Paused Readings 10 Power Save 11 Total Number of Days Statistics Collected 11 Reviewing Day Statistics on the TRID30 7 12 Stopping the TRID30 7 13 Automatically 13 Manually 13 Resetting the Recorde...

Страница 3: ...ice l For any problems with refrigeration units l For the bad quality of the monitored goods if any l For incorrect readings if the device was used with a low battery l For consequential loss Useful L...

Страница 4: ...ter the statistics review and to scroll through the statistical data directly on the display It is also used to place an inspection mark in the data listing Display The display shows at a glance if al...

Страница 5: ...activation type which can be instant consecutive or accumulative1 and a delay time if it is not an instant alarm If an alarm trigger condition requires readings to exceed an upper threshold temperatu...

Страница 6: ...e number of days on the display for which statistical data have been collected see Total Number of Days Statistics Collected on page 11 Parameters influencing specific behaviour of the TRID30 7 are l...

Страница 7: ...time value display is used to show one of the following l a clock with the current time l time remaining to start of logging with a delayed start or l duration for example of an alarm The time indica...

Страница 8: ...uttons To set the clock press and hold the button then press and hold the button Keep holding both buttons together continuously for a period of 8 seconds Press the buttons firmly During this period t...

Страница 9: ...the TRID30 7 starts recording when it reaches 0 00 CURRENT RECORDING TIME The timer can be cancelled and the recorder re set to READY While STARTING and DELAY are shown press and hold the START CLEAR...

Страница 10: ...guration with LogTag Analyzer you can allow a user to clear an alarm on the display This is a useful function for an inspector so new alarms can be recognised easier This display screen shows an exist...

Страница 11: ...not required to be seen frequently such as in transit monitoring applications Pressing any button will re activate the display The power save mode is enabled or disabled when configuring the TRID30 7...

Страница 12: ...m threshold marker l the duration value shows the amount of time recorded above the limit yesterday which was 12 hours 30 minutes l an alarm was triggered indicated by the ALARM as this duration was l...

Страница 13: ...holding the button will get you back to the previous screen rather than advancing one screen Stopping the TRID30 7 Automatically The TRID30 7 automatically stops recording temperature when the maximum...

Страница 14: ...power consumption falls to near zero and the life of the battery approaches its shelf life typically 5 or more years This is useful for conserving battery life when the recorder is not used for extend...

Страница 15: ...configurable lower alarm l ALARM indicator on display linked to alarms Vibration Withstands vibration specification as detailed in EN12830 1999 Shock l Withstands shock specification as detailed in EN...

Страница 16: ...cturer Check the operational temperature range of the battery and make sure it covers the range of the TRID30 7 l Put a new CR2032 cell into the battery compartment as shown the right hand edge of the...