OPERATING ELEMENTS
50
1.
Enter the target quantity for the
should amount
, e.g. 1.00 ml or 1.00 g.
2.
Enter the measured quantity for the
actual quantity
, e.g. 1.15 ml or 0.89 g. Use the
same unit for should amount and actual quantity.
3.
Tap the
Calculate
button to calculate the correction factor.
The button will then change to OK:
Tap
OK
to confirm and apply the correction factor.
Numeric Keypad
Touching a button that requires an entry, opens the following numeric keypad:
Below the display
Min
and
Max
indicate the appropriate input limits. These depend on the
selected pump. If the entered value is outside the limits, it is invalid and will not be accepted.
Note
The correction factor 1.00 must always be set at the beginning of the calibration. To do
this, equal the should amount and actual value and have the correction factor calculated.