1.0 OPERATION:

When you first turn on the UV-V or UV-A/B Radiometer Dosimeter (Meter) by pressing the “On/Off Mode”button once, the system will
automatically test the battery and shut off if the system needs to be re-charged.

1.1 ADAPTER SELECTION

The Meter is designed to be used with Chambers, Conveyors, and Loctite® Led Curing devices. When using certain devices, it is recommended to
use the proper adapters which are purchased separately. The adapter must be installed before taking a reading.

1390323 UV-A/B Radiometer Dosimeter

1265282 UV-V Radiometer Dosimeter

Lamp/LED Device

Adapter

Lamp/LED Device

Adapter

365 LED Wand

1403402

77xx and 77xxHD series LED’

s

1403405

UV Arc Lamps

None Req.

405 Head 976420 & Indigo Head 976418

1403401

375 Head 1369539

1421420

405 Head 1369538 & Indigo Head 1369539

1421420

Visible Arc Lamps

1403403

AssureCure Emitters

1527709

1.2 TAKING A READING: (DATA ACQUISITION MODE)

If the screen is blank, press the “On/Off Mode”button once. The unit will turn on and will display

Screen One

below. The meter will turn off

automatically in 5 minutes if no optical Power is detected.

Screen One

Place the meter

LABEL SIDE DOWN

under the Light Source. After 5 seconds of reading, the display will alternate every 5 seconds between

numeric and graphical displays similar to screens 2 and 3.

Screen Two

shows energy density and irradiance.

Screen three

is a plot of irradiance

versus time.

Screen Two

Screen Three

To stop the display from alternating, press the “On/Off Mode”button once. Press it once again to start the display alternating again. The Meter will
shut itself off automatically after five minutes. To shut the Meter off immediately press the “On/Off Mode”button two times quickly.

Any current reading may be stored for future analysis in the data analysis mode. Please read the Data Acquisition section for details.

2.0

DATA ACQUISITION AND ANALYSIS

The Meter provides a visual display of the status of your LED Lights and lamps as seen on screen three and also has the capability to compare the
current condition of your equipment to that of a user selectable base line or reference condition. The Meter will hold

ONE

irradiance profile in

memory that can be visually compared to a measurement at any time.

2.1. CREATING A BASELINE:

To create a baseline, you will need to take a reading first. Follow instructions in section 1.2. Once you have taken a reading and you have a
profile as shown in

SCREEN 3

, you can create a baseline. To save the current profile as a baseline, press the “ON/OFF Mode”button three times

rapidly. This stores the current profile as your baseline. When you do this, the screen will read all zeros. This profile will remain as your
baseline until you press the “ON/OFF Mode”button three times when a current profile is displayed.