The filter enables to stabilise a weight as long as its variations are smaller than the corresponding
“response time”. It is necessary to set this filter according to the type of application and to the full
scale value set.
FILTER VALUE
Response times
[ms]
Display and serial port refresh
frequency
[Hz]
0
12
300
1
150
100
2
260
50
3
425
25
4 (default)
850
12.5
5
1700
12.5
6
2500
12.5
7
4000
10
8
6000
10
9
7000
5
ANTI PEAK
When the weight is stable, the anti peak filter removes any sudden disturbances with a maximum duration
of 1 second. Confirm the filter on the weight with ENTER and select one of the following options:
-
: anti peak filter enabled (default);
-
: anti peak filter disabled.
ZERO PARAMETERS
RESETTABLE WEIGHT SETTING FOR SMALL WEIGHT CHANGES
(from 0 to full scale; default: 300; considered decimals: 300 – 30.0 – 3.00 – 0.300): this
parameter indicates the maximum weight value resettable by external contact, keypad or serial
protocol.
AUTOMATIC ZERO SETTING AT POWER-ON
(from 0 to 10% of full scale; default: 0): If at switch-on the weight value is lower than the
value set in this parameter, the weight is reset. To disable this function, set 0.
+