LMI LIQUITRON DP5000 Series Скачать руководство пользователя страница 1

Instruction 

5DAAJ

Liquitron

 pH and ORP Wall Mount Controllers

Preamplifier and Cable Assembly

Installation  •  Instructions

 for the DP5000 & DR5000

Fig. A

1. Remove the four screws retaining the
    utility box cover.

2. Install the five wires (orange, blue, black,
    red and green) from the Cable Assembly
    through the PG-9 connector on the right
    side of the controller. Attach the BNC
    connector from the cable to the BNC
    connector on the right side of the controller.

3. Connect wires from the Cable Assembly to
    the controller terminal strip as follows:

Red to

Temp : TB4-8

Black to

Temp : TB4-7

Green to

Terminal Block Ground:
TB1- 1 ,2 or 3

Orange to

Pre-Amp Power +5: TB4-10

Blue to

Pre-Amp Power -5: TB4-9

  WARNING: Disconnect power to the controller

        before opening the front panel!

Neutral

GND

GND

GND

+ 5V

- 5V

NC

TX

RX

Output

PreAmp

Temp

4-20mA

Power

Input

1

2

3

4

5

6

7

8

9

10

11

Output
Pump B

Output
Pump A

Output
Fault

Flow

NO

NO

NO

NO

NC

NC

NO

Low Level

Output

AUX

Remote
on/off

Alarm 1

Alarm 2

Solenoid

Live

120V/230V

7

8

1

2

3

4

5

6

2

3

4

5

Pump B

Pump A

6

7

8

9

10

11

12

1

1

2

3

5

6

7

8

9

4

10

12

Temp

Temp

TB1

TB3

TB4

TB2

Detail A

Replaces same of Rev.C  10/97

1696.D  8/99