Gocator 2300 & 2880 Series
Gocator Web Interface • Scan Setup and Alignment • 75
Mode and Option
Description
Video images are processed internally to produce laser profiles and cross-sectional
measurements.
Surface
Outputs 3D point clouds made up of many laser profiles combined together and
performs surface measurements. The sensor uses various methods to generate a
surface (see on page 100). Part detection can be enabled on a surface to identify
discrete parts (see on page 103).
Uniform Spacing
When this option is enabled, ranges are resampled to a uniform spacing along the X
axis (see on page 46 for more information). The size of the spacing can be set in the
Spacing
tab (see on page 89).
When the option is disabled, the Gocator outputs unprocessed range data. Ranges are
reported in (x,z) coordinate pairs. Post-profiling processing is disabled. Only a subset of
the measurement tools is available.
Disable this option to extract ranges from the Gocator at the highest possible rate.
When the sensor is in Surface mode, this option is not available.
Acquire Intensity
When this option is enabled, an intensity value will be produced for each laser profile
point.
Triggers
A trigger is an event that causes a sensor to take a single picture. Triggers are configured in the
Trigger
panel on the
Scan
page.
When a trigger is processed, the laser is strobed and the camera exposes to produce an image. The
resulting image is processed inside the sensor to yield a profile (range/distance information), which can
then be used for measurement.
The laser and camera inside a sensor can be triggered by one of the following sources:
Trigger Source
Description
Time
Sensors have an internal clock that can be used to generate fixed-frequency triggers.
The external input can be used to enable or disable the time triggers.
Encoder
An encoder can be connected to provide triggers in response to motion. Three encoder
triggering behaviors are supported. These behaviors are set using the
Behavior
setting.
Track Backward
A scan is triggered when the target object moves forward. If the target object moves
backward, it must move forward by at least the distance that the target travelled
backward (this distance backward is "tracked"), plus one encoder spacing, to trigger the
next scan.
Содержание Gocator 2300 Series
Страница 1: ...USER MANUAL Gocator 2300 2880 Series Firmware version 4 3 x xx Document revision D ...
Страница 20: ...Gocator 2300 2880 Series Getting Started System Overview 20 ...
Страница 174: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 174 2D View 3D View Measurement Panel ...
Страница 184: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 184 Measurement Panel ...
Страница 189: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 189 ...
Страница 199: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 199 Measurement Panel ...
Страница 208: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 208 else Output_Set Volume 0 ...
Страница 387: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 387 Gocator 2320 Field of View Measurement Range Dimensions ...
Страница 388: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 388 Envelope Gocator 2330 Field of View Measurement Range ...
Страница 389: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 389 Dimensions Envelope ...
Страница 390: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 390 Gocator 2340 Field of View Measurement Range Dimensions ...
Страница 391: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 391 Envelope Gocator 2342 Field of View Measurement Range ...
Страница 392: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 392 Dimensions Envelope ...
Страница 393: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 393 Gocator 2350 Field of View Measurement Range Dimensions ...
Страница 394: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 394 Envelope ...
Страница 395: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 395 Gocator 2370 Field of View Measurement Range ...
Страница 396: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 396 Dimensions ...
Страница 397: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 397 Envelope ...
Страница 398: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 398 Gocator 2375 Field of View Measurement Range Dimensions ...
Страница 399: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 399 Envelope ...
Страница 400: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 400 Gocator 2380 Field of View Measurement Range ...
Страница 401: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 401 Dimensions ...
Страница 402: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 402 Envelope ...
Страница 405: ...Gocator 2300 2880 Series Specifications Gocator 2880 Sensor 405 Dimensions ...
Страница 406: ...Gocator 2300 2880 Series Specifications Gocator 2880 Sensor 406 Envelope ...