Gocator Line Profile Sensors: User Manual
Gocator Web Interface • 122
To configure X or Z sub-sampling:
1.
Go to the
Scan
page.
2.
Expand the
Sensor
panel by clicking on the panel header or the
button.
3.
Click the button corresponding to the sensor you want to configure.
The button is labeled
Top
,
Bottom
,
Top-Left
, or
Top-Right
, depending on the system.
X and Z sub-sampling is configured separately for each sensor.
4.
Click the
Spacing
tab.
5.
Select an X or Z sub-sampling value.
6.
Save the job in the
Toolbar
by clicking the
Save
button
.
7.
Check that laser profiling is satisfactory.
Spacing Interval
Spacing interval is the spacing between data points in resampled data. (In Profile mode, resampled data
is only produced if the
Uniform Spacing
option in the
Scan Mode
panel is checked.) A larger interval
creates profiles with lower X resolution, reduces CPU usage, and potentially increases the maximum
frame rate. A larger interval also reduces the data output rate. For more information on resampled data,
see
The
Uniform Spacing
option must be checked in the
Scan Mode
panel for the
Spacing Interval
option to be displayed.
You can set the spacing interval either to one of three presets or set a custom value.
To configure the spacing interval:
1.
Go to the
Scan
page.
2.
Choose Profile or Surface mode in the
Scan Mode
panel.
If one of these modes is not selected, you will not be able to configure the spacing interval.
3.
Expand the
Sensor
panel by clicking on the panel header or the
button.
4.
Click the button corresponding to the sensor you want to configure.
The button is labeled
Top
,
Bottom
,
Top-Left
, or
Top-Right
, depending on the system.
Spacing is configured separately for each sensor.
5.
Click the
Spacing
tab.
6.
Do one of the following:
l
Choose
Auto
and move the slider to one of the following values:
Speed
: Uses the lowest X resolution within the active area as the spacing interval. This setting minimizes
CPU usage and data output rate, but the profile has the lowest X resolution (i.e., least detail).
Balanced
: Uses the X resolution at the middle of the active area as the spacing interval. This setting
balances CPU load, data output rate, and X resolution.
Содержание Gocator 2130
Страница 250: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 250...
Страница 262: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 262 2D View 3D View...
Страница 263: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 263 Measurement Panel...
Страница 279: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 279 Measurement Panel...
Страница 285: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 285...
Страница 292: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 292 Measurement Panel...
Страница 298: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 298 2D View 3D View...
Страница 305: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 305 2D View 3D View...
Страница 309: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 309 2D View 3D View...
Страница 586: ...Gocator Line Profile Sensors User Manual Specifications 587 Dimensions Envelope...
Страница 588: ...Gocator Line Profile Sensors User Manual Specifications 589 Envelope...
Страница 590: ...Gocator Line Profile Sensors User Manual Specifications 591 Dimensions Envelope...
Страница 592: ...Gocator Line Profile Sensors User Manual Specifications 593 Dimensions Envelope...
Страница 594: ...Gocator Line Profile Sensors User Manual Specifications 595 Dimensions...
Страница 595: ...Gocator Line Profile Sensors User Manual Specifications 596 Envelope...
Страница 597: ...Gocator Line Profile Sensors User Manual Specifications 598 Dimensions...
Страница 598: ...Gocator Line Profile Sensors User Manual Specifications 599 Envelope...
Страница 600: ...Gocator Line Profile Sensors User Manual Specifications 601 Dimensions...
Страница 601: ...Gocator Line Profile Sensors User Manual Specifications 602 Envelope...
Страница 603: ...Gocator Line Profile Sensors User Manual Specifications 604 Dimensions...
Страница 604: ...Gocator Line Profile Sensors User Manual Specifications 605 Envelope...
Страница 608: ...Gocator Line Profile Sensors User Manual Specifications 609 Dimensions...
Страница 609: ...Gocator Line Profile Sensors User Manual Specifications 610 Envelope...
Страница 611: ...Gocator Line Profile Sensors User Manual Specifications 612 Dimensions...
Страница 612: ...Gocator Line Profile Sensors User Manual Specifications 613 Envelope...
Страница 615: ...Gocator Line Profile Sensors User Manual Specifications 616 Dimensions...
Страница 616: ...Gocator Line Profile Sensors User Manual Specifications 617 Envelope...