Gocator Line Profile Sensors: User Manual
How Gocator Works • 59
For more information, see the
or the
, and the
Spacing (Data Resampling)
Data produced in Profile mode is available in two formats: with and without uniform spacing. Uniform
spacing is enabled in the
Scan Mode
panel, on the
Scan
page.
When uniform spacing is enabled, the ranges that make up a profile are resampled so that the spacing is
uniform along the laser line (X axis). The resampling divides the X axis into fixed size "bins." Profile points
that fall into the same bin are combined into a single range value (Z). The size of the spacing interval is
set under the
Spacing
tab in the
Sensor
panel on
Scan
page.
Resampling to uniform spacing reduces the complexity for downstream algorithms to process the profile
data from the Gocator, but places a higher processing load on the sensor's CPU.
When uniform spacing is not enabled, no processing is required on the sensor. This frees up processing
resources in the Gocator, but usually requires more complicated processing on the client side. Ranges in
this case are reported in (X, Z) coordinate pairs.
Most built-in measurement tools in the Gocator in Profile mode operate on profiles with uniform
spacing. A limited number of tools can operate on profiles without uniform spacing. For more
information on the profile tools, see
A drawback of uniform spacing is that if sensors are angled to scan the sides of a target, data on the
"verticals" is lost because points falling in the same "bin" are combined. When
Uniform Spacing
is
disabled, however, all points are preserved on the sides. In this case, the data can be processed by the
subset of tools that work on profiles without uniform spacing. Alternatively, the data can be processed
externally using the SDK.
When uniform spacing is enabled, in the Ethernet output, only the range values (Z) are reported
and the X positions can be reconstructed through the array index at the receiving end (the
client). For more information on Ethernet output, see
For information on enabling uniform spacing, see
Содержание Gocator 2130
Страница 250: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 250...
Страница 262: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 262 2D View 3D View...
Страница 263: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 263 Measurement Panel...
Страница 279: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 279 Measurement Panel...
Страница 285: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 285...
Страница 292: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 292 Measurement Panel...
Страница 298: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 298 2D View 3D View...
Страница 305: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 305 2D View 3D View...
Страница 309: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 309 2D View 3D View...
Страница 586: ...Gocator Line Profile Sensors User Manual Specifications 587 Dimensions Envelope...
Страница 588: ...Gocator Line Profile Sensors User Manual Specifications 589 Envelope...
Страница 590: ...Gocator Line Profile Sensors User Manual Specifications 591 Dimensions Envelope...
Страница 592: ...Gocator Line Profile Sensors User Manual Specifications 593 Dimensions Envelope...
Страница 594: ...Gocator Line Profile Sensors User Manual Specifications 595 Dimensions...
Страница 595: ...Gocator Line Profile Sensors User Manual Specifications 596 Envelope...
Страница 597: ...Gocator Line Profile Sensors User Manual Specifications 598 Dimensions...
Страница 598: ...Gocator Line Profile Sensors User Manual Specifications 599 Envelope...
Страница 600: ...Gocator Line Profile Sensors User Manual Specifications 601 Dimensions...
Страница 601: ...Gocator Line Profile Sensors User Manual Specifications 602 Envelope...
Страница 603: ...Gocator Line Profile Sensors User Manual Specifications 604 Dimensions...
Страница 604: ...Gocator Line Profile Sensors User Manual Specifications 605 Envelope...
Страница 608: ...Gocator Line Profile Sensors User Manual Specifications 609 Dimensions...
Страница 609: ...Gocator Line Profile Sensors User Manual Specifications 610 Envelope...
Страница 611: ...Gocator Line Profile Sensors User Manual Specifications 612 Dimensions...
Страница 612: ...Gocator Line Profile Sensors User Manual Specifications 613 Envelope...
Страница 615: ...Gocator Line Profile Sensors User Manual Specifications 616 Dimensions...
Страница 616: ...Gocator Line Profile Sensors User Manual Specifications 617 Envelope...