Gocator Line Profile Sensors: User Manual
Gocator Web Interface • 271
Parameter
Description
Absolute Threshold
When
Use Intensity
is disabled, the setting specifies the minimum
height
difference
between points on a path profile for that step to be considered for an edge point.
The setting can be used to exclude smaller steps on a part that should not be
considered for an edge, or to exclude height differences caused by noise. When used in
conjunction with
Relative Threshold
,
Absolute Threshold
is typically set to a small
value, greater than the general surface roughness.
In the image below, when
Step Threshold
is left at the default of 0, all steps are
included as possible candidates for an edge, and will be used to fit an edge line. The
resulting edge line is angled upward to the left.
When
Step Threshold
is set to 3 with the same data (see image below), steps going from
the yellow to pink regions (roughly 1.37 mm) and from the blue to yellow regions (roughly
Содержание Gocator 2130
Страница 250: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 250...
Страница 262: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 262 2D View 3D View...
Страница 263: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 263 Measurement Panel...
Страница 279: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 279 Measurement Panel...
Страница 285: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 285...
Страница 292: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 292 Measurement Panel...
Страница 298: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 298 2D View 3D View...
Страница 305: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 305 2D View 3D View...
Страница 309: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 309 2D View 3D View...
Страница 586: ...Gocator Line Profile Sensors User Manual Specifications 587 Dimensions Envelope...
Страница 588: ...Gocator Line Profile Sensors User Manual Specifications 589 Envelope...
Страница 590: ...Gocator Line Profile Sensors User Manual Specifications 591 Dimensions Envelope...
Страница 592: ...Gocator Line Profile Sensors User Manual Specifications 593 Dimensions Envelope...
Страница 594: ...Gocator Line Profile Sensors User Manual Specifications 595 Dimensions...
Страница 595: ...Gocator Line Profile Sensors User Manual Specifications 596 Envelope...
Страница 597: ...Gocator Line Profile Sensors User Manual Specifications 598 Dimensions...
Страница 598: ...Gocator Line Profile Sensors User Manual Specifications 599 Envelope...
Страница 600: ...Gocator Line Profile Sensors User Manual Specifications 601 Dimensions...
Страница 601: ...Gocator Line Profile Sensors User Manual Specifications 602 Envelope...
Страница 603: ...Gocator Line Profile Sensors User Manual Specifications 604 Dimensions...
Страница 604: ...Gocator Line Profile Sensors User Manual Specifications 605 Envelope...
Страница 608: ...Gocator Line Profile Sensors User Manual Specifications 609 Dimensions...
Страница 609: ...Gocator Line Profile Sensors User Manual Specifications 610 Envelope...
Страница 611: ...Gocator Line Profile Sensors User Manual Specifications 612 Dimensions...
Страница 612: ...Gocator Line Profile Sensors User Manual Specifications 613 Envelope...
Страница 615: ...Gocator Line Profile Sensors User Manual Specifications 616 Dimensions...
Страница 616: ...Gocator Line Profile Sensors User Manual Specifications 617 Envelope...