Gocator Line Profile Sensors: User Manual
Gocator Web Interface • 163
Setting
Description
Match Algorithm
Determines which algorithm the sensor will use to attempt a match. Set this to
Edge
for edge detection.
Image Type
Determines what kind of data the Gocator will use to detect edges and therefore
for part matching. Choose this setting based on the kinds of features that will be
used for part matching:
Heightmap
: Surface elevation information of the scanned part will be used to
determine edges. This setting is most commonly used.
Intensity
: Intensity data (how light or dark areas of a scanned part are) will be
used to determine edges. Use this setting if the main distinguishing marks are
printed text or patterns on the parts. The
Acquire Intensity
option must be
checked in the
Scan Mode
panel on the
Scan
page for this option to be
available.
Z Angle
Corrects the orientation of the model to accurately match typical orientation and
simplify measurements.
Target Sensitivity (
Target
Matching
tab)
Controls the threshold at which an edge point is detected on the target's
heightmap or intensity image. (The "target" is any part that is matched to the
model and which will subsequently be measured if the match is accepted.)
Setting
Target Sensitivity
higher results in more edge points. Setting it lower
results in fewer edge points and results in higher performance. Use this setting to
exclude noise from the detected edges and to make sure distinguishing features
are properly detected.
The level of this setting should generally be similar to the level of
Model
Sensitivity
.
Model Sensitivity
(
Model Editing
tab)
Controls the threshold at which an edge point is detected on the heightmap or
intensity image used to create the model. Setting
Model Sensitivity
higher
results in more edge points. Setting it lower results in fewer edge points and
results in higher performance. Use this setting to exclude noise from the detected
edges and to make sure distinguishing features are properly detected.
The level of this setting should generally be similar to the level of
Target
Sensitivity
.
Changing this setting causes the edge detection algorithm to run again at the new
threshold. If you have edited edge points manually (removing them selectively),
those changes will be lost. See
on page 161 for more
information.
Edge Points
(
Model Editing
tab)
The
Edit
button lets you selectively remove edge point that are detected by the
edge detection algorithm at the current
on page 161 for more information.
Acceptance Criteria
Determines the minimum quality level of the match as a percentage value.
Part rejected: Quality result is less than Min
To run part matching, simply make sure that the
Enabled
option is checked on the
Part Matching
panel when the Gocator is running. Any measurements that are added and configured on the
Measure
Содержание Gocator 2130
Страница 250: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 250...
Страница 262: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 262 2D View 3D View...
Страница 263: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 263 Measurement Panel...
Страница 279: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 279 Measurement Panel...
Страница 285: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 285...
Страница 292: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 292 Measurement Panel...
Страница 298: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 298 2D View 3D View...
Страница 305: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 305 2D View 3D View...
Страница 309: ...Gocator Line Profile Sensors User Manual Gocator Web Interface 309 2D View 3D View...
Страница 586: ...Gocator Line Profile Sensors User Manual Specifications 587 Dimensions Envelope...
Страница 588: ...Gocator Line Profile Sensors User Manual Specifications 589 Envelope...
Страница 590: ...Gocator Line Profile Sensors User Manual Specifications 591 Dimensions Envelope...
Страница 592: ...Gocator Line Profile Sensors User Manual Specifications 593 Dimensions Envelope...
Страница 594: ...Gocator Line Profile Sensors User Manual Specifications 595 Dimensions...
Страница 595: ...Gocator Line Profile Sensors User Manual Specifications 596 Envelope...
Страница 597: ...Gocator Line Profile Sensors User Manual Specifications 598 Dimensions...
Страница 598: ...Gocator Line Profile Sensors User Manual Specifications 599 Envelope...
Страница 600: ...Gocator Line Profile Sensors User Manual Specifications 601 Dimensions...
Страница 601: ...Gocator Line Profile Sensors User Manual Specifications 602 Envelope...
Страница 603: ...Gocator Line Profile Sensors User Manual Specifications 604 Dimensions...
Страница 604: ...Gocator Line Profile Sensors User Manual Specifications 605 Envelope...
Страница 608: ...Gocator Line Profile Sensors User Manual Specifications 609 Dimensions...
Страница 609: ...Gocator Line Profile Sensors User Manual Specifications 610 Envelope...
Страница 611: ...Gocator Line Profile Sensors User Manual Specifications 612 Dimensions...
Страница 612: ...Gocator Line Profile Sensors User Manual Specifications 613 Envelope...
Страница 615: ...Gocator Line Profile Sensors User Manual Specifications 616 Dimensions...
Страница 616: ...Gocator Line Profile Sensors User Manual Specifications 617 Envelope...