22
Error Code
Sc
a
le
r
Pol
is
he
r
Diagnosis instructions
Cause code
Repair / Remedy instructions
working handpiece the handpiece is faulty.
and submit to LM fax +358 2 454 6444. After obtaining an
RMA number from LM, return the faulty handpiece for
replacement.
2. If there is no Light with the working
handpiece, the tubing is probably damaged.
C-002 Faulty tubing
Level 1 Service Providers:
Fill in
and submit to LM fax +358 2 454 6444. After obtaining an
RMA number from LM, return the entire device with
accessories and the RMA number to LM for repair.
Level 2 Service Providers:
Replace tubing. See repair instructions in
Fill in
and submit to LM fax +358 2 454 6444. After obtaining an
RMA number from LM, return the faulty hose for replacement.
3. If above actions do not solve the problem,
request authorization to return the device to
LM-Instruments
C-007 Unidentified loss
of handpiece light
Submit
to LM fax +358 2 454 6444. After obtaining an RMA number
from LM, return the entire device with accessories for repair.
E-S04 No/low water flow
X
1. Double check that:
-
Water control ring on handpiece is
fully open
-
the tip is removed (or new)
C-017 Stuck waterline
in handpiece
Use the LM-ProPower Fixer tool to open up the stuck water
line.