Page 94 of 130
Operating instructions
4point light 10W (V 4.58)
To
select an individual provider
, proceed as follows:
1. Switch to the
menu item „Misc“ „Provider“ and request a valid provider list
➔
8.11.1 Provider, pages 99
2.
Switch to „Settings“ „Communication“ „Provider“
3. Enter the provider ID of the desired mobile operator by using the numerical
keys
Provider ID:
26207
Fig. 8.105: Enter provider ID
4. Confirm input with [RET]
By default,
no provider is selected (Provider ID: 0). In this case
the instrument automatically
chooses the strongest available access point.
8.10.3 Service
Under the menu item “Service” the factory settings can be adjusted. A code number is
required for access.
8.10.4 Acquisition
Under the menu item “Acquisition” multiple parameters for data acquisition can be set
and country specific frequencies can be selected.
! NOTICE
Noise signal at 10 Hz measurement frequency
In Europe or other countries with 50 Hz mains frequency measurements at 10 Hz
result in strong interference signals. These signals interfere with the measurement.
→
Never use 10 Hz in Europe or in any other countries with 50 Hz mains
frequency.