LN0405E2107B0 

        

         

   

       

Vaporizer Unit Instruction Manual 

 

                        VU-0430N Series 

 

Safety Precautions 

 

   

  Incorrect handling can cause serious injury or death. 

 

(1) 

Before connecting the fittings, check that no damage or defects are found 
on the fittings. Make connections properly and make sure that a leak test is 
conducted before actual operation to prevent liquid and gas from leaking 
into the atmosphere. 

(2) 

DO NOT apply any corrosive fluid to materials exposed to liquid and gas. 
Corrosion may cause liquid or gas to leak into the atmosphere. Please 
confirm the physical properties of liquid or gas before using. 

(3) 

This device is not designed as an explosion-proof structure. DO NOT use 
this device in a place where explosion-proof structures are required. Doing 
so may cause fire or explosion.   

(4) 

Prepare temperature controller unit when operating Vaporizer/Heat Exchanger 
and do not set the temperature over than maximum operating temperature. 
Wring temperature setting may cause fire or destruction of the device. It is 
recommended to add abnormal overheating detector if necessary. 

(5) 

This device must be earthed before use. Otherwise, there is the rink of 
electric shock.   

(6) 

Thermal switch is equipped in Vaporizer/Heat Exchanger to prevent 
overheating. However, the operating temperature of thermal switch would 
vary due to operating conditions and ambient temperature.   

(7) 

Attach/remove connector and terminals, please make sure that power 
supply turning off. It may cause fire or shock hazard.   

 

Incorrect handling can cause medium or

 

slight injury or 

may cause

 

damage to, or loss of, facilities or equipment. 

 

(1) 

Observe the precautions listed in the WARNING (above). 

(2) 

Not using the voltage can cause fire, damage to sensors or malfunction or 
cause electric shock. 

(3) 

This device is not designed to be waterproof. DO NOT locate this product 
outdoors or in a place where it may be splashed with water. Doing so may 
cause fire, trouble or malfunction of this product. 

(4) 

DO NOT modify this product. It may cause fire or other problems. 

(5) 

A  warm-up  period  of  60  minutes  is  recommended  after  reaching  the  set 
temperature. Otherwise, the output gas temperature will be low. 

(6) 

This  device  is  a  precious  device,  please  handle  it  carefully.  Dropping 
down  or  handling  it  carelessly  will  cause  damage.  Please  use  assist 
instrument while moving or setting the device. 

(7) 

Depends  on  the  character  of  the  liquid  and  process  condition,  the 
temperature  of  tube,  carrier  gas,  and  the  VU,  also  the  flow  rate  of  carrier 
gas should be set properly. 

(8) 

Please  use  Helium  (He)  to  pressure  liquid.  It  easily  lead  to  bubbles  if  use 
N2, Ar, O2, etc, that have higher than Helium (He) solubility to pressure. 

(9) 

Regular maintenance is recommended for the steady use of this product. 
This product uses a seal

 

material made from the Kalrez®, in order to 

reduce leakage, Changing the seals annually is recommended. The 
maintenance cycle changes according to precursor type and operating 
conditions, so please consult about the maintenance cycle for the steady use 
of this device. 

 

1. Introduction 

This  instruction  manual  explains  the  basic  operation  of  Vaporizer  unit

 

VU-0430N  Series  (Hereafter  called  “VU”).  Please  read

 

through  this  manual 

carefully to become familiar with the features of this device.   
 

2. Features 

The VU has the following features. 

(1) 

By using the VU in combination with the liquid mass flow meter 
(Hereafter called “LM”) and the gas mass flow controller (Hereafter 
called “MFC”), precise flow control and effective vaporization can be 
assured. 

(2) 

Maximum operating temperature of  200℃. 

(3) 

Compact size 

 
 
 

3. Specification/ Dimensions 

   

(1) Specification 

Product name  

Vaporizer 

Model name   

VU-0430N 

Rated power 
supply voltage 
and heater 
capacity 

Vaporizer model: D01 

Inner AC 120V 80W 

Outer AC 120V 100W 

Vaporizer model: D02 

Inner AC 240V 225W  

Outer AC 240V 400W  

Thermal switch (Note1) 

230±10°C

 

OPEN  

The rma l s w itc h c onne c tion  me thod  

Outer block: 1piece, Inner block: 1piece 

Direct connection (125V/15A, 250V/10A) 

Operating primary pressure  

150 to 300 kPa(G) 

Operating secondary  pressure  

Less than 25 kPa(G) 

Pressure  Limit  (Gauge pressure)  

1MPa(G) 

Leak integrity  (External Leak)  

Less than 5.0×10

-10 

Pa∙m

3

/sec (He) 

Fitt ing s  

Precursor inlet 

3.2mm (1/8”) VCR Male 

Carrier gas inlet 

6.35mm (1/4”) VCR Male 

Gas outle t 

6.35mm (1/4”) VCR Female 

Th ermo cou ple  fo r t emp erat ure  co ntroll er  

K Type × 2pieces 

Reco mmen d ed t emperat ure  co ntroll er  

PID control type 

Wetted  materials  

StainlessSteel316L, Au, Ni-Co,   

Polyimide or PEEK 

Mounting  direction  

Piping line: horizontal. 

Power supply connector: vertical and up. 

(Please confirm the outline dimension drawing) 

Standard

 

accessories  

Thermocouple 
connector 

CMP01-K(RKC ) × 2pieces  

Heater/Thermal switch 
connector 

SRCN6A16-7S (JAE) × 1piece 

LM-VU connect 
cable  

CC-LV-3-3M× 1piece  

 

 

Note1) The temperature of thermal switch may be different from the temperature of a point of 

     

 

 

controlling of VU. 

                         

 

 

 

(2) Dimensions   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fig1.   Exte rnal  dimens ions   and  com ponents  

 

 