Lintec MC-3000S Series Скачать руководство пользователя страница 1

 

LN0404E2109B0   

         

 

Low differential pressure Mass Flow Controller

 

Instruction Manual 

   

MC-3000S Series

 

  

Safety Precautions 

 

 
Incorrect handling may cause death or injury

 

 

(1)

 

Before connecting the fittings, check that no damage or defects are found on 
the  fittings.  Make  connections  properly  and  make  sure  that  a  leak  test  is 
conducted  before  actual  operation  to  prevent  fluid  from  leaking  into  the 

atmosphere (Hereafter, the measured fluid is called “gas” or “fluid”).

 

(2)

 

DO NOT

 apply any fluids corrosive to materials exposed to gas. Corrosion may cause 

fluid to leak into the atmosphere. Check the gas type to be used in advance.

 

(3)

 

This device is not designed as an explosion proof structure. 

DO NOT

 use this 

device in a place where explosion-proof structures are required. Doing so may 
cause fire or explosion.

 

 

Incorrect handling may lead to medium or slight injury or may 
cause damage to, or loss of, facilities or equipment

 

 

(1)

 

Observe the precautions listed in the WARNING (above).

 

(2)

 

Strictly  observe  the  electrical  specifications.  Not  doing  so  may  cause  fire, 
damage to sensors or malfunction.

 

(3)

 

This  device  is  not  designed  to  be  waterproof. 

DO  NOT

  locate  this  device 

outdoors or in a place where it may be splashed with water. Doing so may cause 
fire, trouble, or malfunction of the device.

 

(4)

 

DO NOT

 modify this device. It may cause fire or other problems. 

(5)

 

While a power supply is applied to the device, +15VDC and -15VDC must be 

applied  simultaneously.  If  only  +15VDC  or  -15VDC  is  applied,  electronic 

circuits will become unstable and it may cause a malfunction.

 

(6)

 

This  device  is  not  designed  to  handle  hot  swap.  Please  avoid  attaching  and 
removing the power supply connector and interface connector with the power 
switched  on.  Attachment  and/or  removal  with  the  power  on  may  result  in 
failure of the device.

 

(7)

 

This device is a precious device, please handle it carefully. Dropping down or 
handling  it  carelessly  will  cause  damage.  Please  use  assist  instrument  while 
moving or setting the device.

 

(8)

 

Regular maintenance is recommended for steady use of this device 
(Recommended proofreading frequency is once a year). 

 

 

1. Introduction

 

   

This manual explains basic operation of the MC-3000S series (Hereafter, it is called 
"MFC").  Please  read  through  this  manual  and  other  separate  volumes  (Digital 
Interface  Manual,  Special  Function  Manual,  Command  Chart)  carefully  to 
familiarize yourself with the features of this device. 

 

2. Summary 

   

The MFC is high performance mass flow controller for gas using thermal flow 
sensor and high response piezoelectric valve. In the MFC model, MO type and 
MC type are available depending on the kind of valve operation mode.

 

 

3. Features 

The MFC is the following features. 

(1)

 

LINTEC’s proprietary ambient temperature compensation type flow sensor is carried.

 

 

  ·

 

Low sensor temperature, long-time steady monitor, and few problems such 

       

           

as degradation of gas to be monitored. 

·

 

The effect of ambient temperature is less because the sensor temperature is 

 

   

kept at the ambient temperature.   

     

         

·

 

Steady temperature distribution of sensor and high-speed response. 

(2)

 

Digital computing system with microprocessor and high-resolution A/D, D/A 
converter is carried. 

 

· High functionality. 
·

   

By setting of device number (address), multiple devices can be controlled   

 

   

through a single interface. 

· Many additional are provided as standard, e.g., totalizer function, zero adjust, 
 

 

ramping function. 

(3)  Low differential pressure device.

 

(4)  Universal digital Interface (RS-485). 
(5)

 

   

Small dead space structure using a diaphragm valve. 

(6)

 

Superior  corrosion  resistant  sealing materials  make maintenance easy.  Metal 
seals (Au). 

(7)

 

Particle-free structure.

   

(8)

 

By using a metal case and various types of filters, steady operation can achieve 
even in an environment of high-frequency noise and stationary magnetic field.

 

(9)

 

Based on RoHS. 

 

 

4. Structure 

   

The  MFC  consists  of  sensors,  bypass,  valves,  and  a  microcomputer  for  signal 
processing. A digital PID feedback control system controls the valve action so that 
flow rate output from the sensor agrees with flow rate setting value. 

 
 
 
 
 
 
 
 

5. Specification / Dimensions 

(1) Specification 

Name 

Low differential pressure Mass Flow Controller 

Model 

MC-3102S 

Standard flow rate in nitrogen 

2 to 10SCCM 

~20SCCM 

~30SCCM 

Valve operation mode 

Normally open valve / Normally closed valve 

Internal surface treatment 

No treatment / Precision polishing 

Minimum controllable flow rate 

2%F.S. 

Analog flow rate output signal 

Proportional to flow rate 0 to 5VDC 

Analog flow rate setting signal 

Accuracy 

±1.0%F.S. 

Repeatability 

±0.2%F.S. 

Response time 

2 second (Typical) 

Operating differential pressure 

8.00×10

2

Pa to 

1.33×10

5

Pa 

1.07×10

3

Pa to 

1.33×10

5

Pa 

1.33×10

3

Pa to 

1.33×10

5

Pa 

Maximum operating pressure  

1.33×10

5

Pa 

Withstanding pressure 

1MPa(G) 

Operating temperature & 
Humidity 

5 to 50 ºC·0 to 80%RH (Condensation should be avoided) 

(Accuracy insurance 15 to 35℃) 

Temperature coefficient 
(Standard Temp.25 ºC)  

Zero 

±0.02%F.S./ºC 

Span 

±0.02%F.S./ ºC 

±0.04%F.S./ ºC 

±0.08%F.S./ ºC 

Storage temperature & Humidity  

0 to 60 ºC·0 to 80%RH (Condensation should be avoided) 

Leak integrity 

Less than 1×10

-11

Pa · m

3

/sec (He) 

Mounting direction 

Specified upon placement of order 

(Please see below (2) Mounting direction for details.) 

Wetted materials 

Stainless steel 304, Stainless steel 316L, PCTFE, Au 

Seal Materials 

Au 

Actuator 

Piezoelectric actuator 

Fittings 

6.35VCR (106, 124)        6.35UJR (106, 124) 

1.125W-Seal (79.8mm, 92mm)

 

1.5W-Seal (79.8mm, 92mm) 

Power supply 

+15V DC±3%: 100mA      -15VDC±3%: 50mA 

Digital interface 

RS-485 (2-wire) 

Weight 

Approx. 1kg 

·

  Connect the MFC to the frame ground

 

(2) Mounting direction 

 

(3) Dimensions 

 

6. Ordering information 

 

M C - 3 1 0 2 S   –   N C   –   4 J R 1   J V 0 A 0   –   d d   –   e e   –   f f  

       

           

[1]

   

       

[2]

 

     

[3]

   

         

[4]

 

     

                                         

   

[5]

   

           

[6]

   

           

[7]

 

 

[1] Series model

 

MC: Mass Flow Controller    MC Series 

[2] Valve mode 

   

NO: Normally open (No treatment), MO: Normally open (Precision polishing) 

   

NC: Normally close (No treatment), MC: Normally close (Precision polishing) 

[3] Fitting 

   

4JR1:6.35UJR106, 

 

4JR2:6.35UJR124,    4VR1:6.35VCR106,    4VR2:6.35VCR124,    4SWL:6.35SWL 

 

      

※ Please consult for more information. 

[4] Optional 

   

J:

     

Valve is closed at below 2% setting 

    Mounting direction

 

  V: HBD,

 

W: VID,

 

X: VIU 

   

※ Default setting is labeled “VA0A0”. Please consult for more information. 

[5] Gas type 
[6] Full scale flow rate 
[7] Flow rate unit

 

SCCM (0 ºC standard), SLM (0 ºC standard) 