Gloss Meter 

User Manual V3.1 

 

 

 

Please read the manual carefully before you use. 

 

Содержание LS191

Страница 1: ...Shenzhen Linshang Technology Co Ltd 1 4 2019 Page 1 of 16 Gloss Meter User Manual V3 1 Please read the manual carefully before you use ...

Страница 2: ...ibration The working temperature of the meter should fall between 10 40 and the relative humidity should be less than 85 without congealed hoarfrost The storage temperature of the meter should fall between 10 60 The meter should be kept away from heat sources and corrosive agents and stored in a dry place The meter should be used without strong electromagnetic interference The meter provides one y...

Страница 3: ... 4 2019 Page 3 of 16 Table of Contents I About the meter 4 II Structure 6 III Technical parameters 6 IV Operation 7 1 Powering on 7 2 Auto Calibration 9 3 Measurement 12 4 Powering off 13 V Maintenance 14 VI Factory list 15 VII Service 16 ...

Страница 4: ...r class one working meters Table 1 Verification performance of the gloss meter LS191 LS192 features To provide real time measurement and put and measure services without pressing buttons It displays the measurement data on a real time basis Equipped with intelligent statistics function The number of measurements the maximum value the minimum value the average value and the standard deviation are d...

Страница 5: ...t Aviation aluminum housing and Calibration Holder with super good quality small design and being easy to carry Built in rechargeable lithium battery with Ultra low power consumption LS191 LS192 is widely used for Gloss measurement of materials and products in the fields of paint coatings decorative materials building materials plastic materials ceramic products stone products wood products leathe...

Страница 6: ...5 9 15 Measurement range GU 0 200 0 1000 Resolution GU 0 1 0 1 Repeatability GU 0 100 0 2 100 200 0 2 0 100 0 2 100 1000 0 2 Reproducibility GU 0 100 0 5 100 200 0 5 0 100 0 5 100 1000 0 5 Zero error GU 0 1 0 1 Error of indication GU 0 100 1 5 100 200 1 5 0 100 1 5 100 1000 1 5 Working temperature 10 40 10 40 Storage temperature 10 60 10 60 Relative humidity Less than 85 no dew Less than 85 no dew...

Страница 7: ...r the status of powering off press the button of Power to power on the meter Normally the meter will successively enter into start interface warm up interface autodiagnosis interface auto calibration interface and measurement interface The whole process will last about 10 seconds After entering into the measurement interface the users can make the measurement Start interface Warm up interface warm...

Страница 8: ...ystem errors improper operation and environmental changes Once the autodiagnosis result is normal the system will display OK with the meter entering into the auto calibration interface Autodiagnosis function to check whether the calibration standard is clean If the standard value of the calibration standard falls out of qualified range the meter will display Please clean standard ...

Страница 9: ... restart the meter for measurement after confirming that the meter temperature tends to be equal to the environmental temperature and that lens in the measuring port and the standard on the Calibration Holder are free from congealed hoarfrost After ruling out the above problems place again the meter into the Calibration Holder and the meter will run again the autodiagnosis If the troubles are remo...

Страница 10: ... Key and the meter will run auto calibration After calibration it will enter into measurement mode and the measuring value is the same as that shown by the label on the right side of the meter If the no key pressed before the number reaches 0 the meter will directly enter into the measuring status without calibration operation If the system displays Please call service after calibration the reason...

Страница 11: ...ating Use the special lens cloth as normal paper towels and cloths cannot be used to clean the standard provided by our company when you clean the standard Please use the cloth after cleaning if the cloth is dirty 2 If the meter does not start in the Calibration Holder it will not run autodiagnosis and auto calibration but directly enter into measurement status 3 The standard value 94GU above is d...

Страница 12: ... to 1 after reaching the counting times and maximum value the minimum value the average value and the standard deviation will reset Every time you press the button the meter will automatically count the maximum value the minimum value the average value and the standard deviation The default counting times is 10 and the adjustable range is 1 99 See the User Manual of Gloss Meter for counting times ...

Страница 13: ... meter will automatically power off 3 minutes after the users carry out no operation Operation notes 1 Please ensure that the calibration standard is clean and the meter is pressed into the Calibration Holder before powering on so that the meter can run a successful autodiagnosis If the meter starts in a place other than the Calibration Holder the meter will not run auto diagnosis and auto calibra...

Страница 14: ...er for measurement after confirming that the meter temperature tends to be equal to the environmental temperature and that lens in the measuring port and the standard on the Calibration Holder are free from congealed hoarfrost 6 If there are multiple meters put the meter on the Calibration Holder corresponding to the serial number of the meter for calibration 7 When not in use place the meter on t...

Страница 15: ...cessively discharged and damaged 5 Avoid to contact the corrosive agents and keep away from high temperature and high humidity environment 6 Put the meter in the specialized package after powering off and keep it in a safe place 7 The calibration period is suggested to be one year and the company provides calibration service VI Factory list Gloss meter 1 Calibration Holder with the standard 1 USB ...

Страница 16: ...e whole meter to the company for maintenance Provide users with spare parts and lifelong maintenance services Provide the users with the meter inspection service for free Free technical support for long term Manufacturer Shenzhen Linshang Technology Co Ltd Website www linshangtech com Service hotline 086 755 86263411 Email sales21 linshangtech com ...