Shenzhen Linshang Technology Co., Ltd.
1-3-2019 Page 1 of 3
LS110 split transmission meter
User Manual V8.1
LS110 is used for the transmittance testing of glass, plated film, solar film,
insulating
glass
, filmed glass etc.
LS110 consist of two parts: PART
Ⅰ
(main part) for measuring and displaying
functions; PART
Ⅱ
(the auxiliary part) for light source and laser, the laser is used for two
part alignment.
I. LS110 Parameters
1.Dimension: 130mm × 70mm × 84mm (L × W ×
H)
2.Weight: 630g
3.Resolution: 0.1% for measured value<10%
and 1% for other ranges
4.Accuracy: better than ±2% (Colorless and
transparent material)
5.Test sample size: > 40mm×40mm
6.Visible light wavelength: 380nm-760nm, CIE
function of photopic vision
7.Power supply: built-in lithium rechargeable battery
II
.
Key Operation
1
.
Power on/off
Short press "Power" key to power the instrument on/off
,
main and auxiliary part
are
required to perform the operation of power on/off.
The meter is shut down automatically for no action within 15 minutes.
2. “HOLD” Key of the main part
Under the measurement mode, short press "HOLD" to hold the data on LCD, up right
corner of LCD will display the icon "HOLD" .
Short press "HOLD" again to release the hold mode and the icon "HOLD" disappears.
Long press "HOLD" for 3 seconds to 3 value display mode. Long press "HOLD" for 3
seconds again to 1 value display mode. In fact, the instrument is composed of three sets
of independent transmittance meter. 1 value mode is the mean values of three 3 value
mode.
3.”Laser”key of the auxiliary part
Short press "Laser" key to power the laser on/off
,
III. Instrument operation
1. SELF-TEST
Attracting the main and auxiliary part together, press down “POWER" key to power
them respectively and align the upper and lower laser points with the alignment bull’s
eyes simultaneously. Wait for the completion of “SELF-TEST” and enter the
measurement interface(100% displayed on the LCD). The following reasons for failed
“SELF-TEST”(“Please refer to the manual” display on the LCD) :