                                               Shenzhen Linshang Technology Co., Ltd. 

1-3-2019    Page 1 of 3 

LS110 split transmission meter 

User Manual V8.1 

LS110 is used for the transmittance testing of glass, plated film, solar film, 

insulating 

glass

, filmed glass etc. 

LS110 consist of two parts: PART

(main part) for measuring and displaying 

functions; PART

  (the auxiliary part) for light source and laser, the laser is used for two 

part alignment. 

I. LS110 Parameters

 

1.Dimension: 130mm × 70mm × 84mm (L × W × 

H) 

2.Weight: 630g 
3.Resolution: 0.1% for measured value<10% 

and 1% for other ranges 

4.Accuracy: better than ±2% (Colorless and 

transparent material) 

5.Test sample size: > 40mm×40mm 
6.Visible light wavelength: 380nm-760nm, CIE 

function of photopic vision 

7.Power supply: built-in lithium rechargeable battery 

II

Key Operation 

1

Power on/off 

Short press "Power" key to power the instrument on/off

main and auxiliary part 

are 

required to perform the operation of power on/off. 

The meter is shut down automatically for no action within 15 minutes.

 

2. “HOLD” Key of the main part 

Under the measurement mode, short press "HOLD" to hold the data on LCD, up right 

corner of LCD will display the icon "HOLD" . 

Short press "HOLD" again to release the hold mode and the icon "HOLD" disappears. 
Long press "HOLD" for 3 seconds to 3 value display mode. Long press "HOLD" for 3 

seconds again to 1 value display mode. In fact, the instrument is composed of three sets 
of independent transmittance meter. 1 value mode is the mean values of three 3 value 
mode.  

3.”Laser”key of the auxiliary part 
     

Short press "Laser" key to power the laser on/off

 

III. Instrument operation 

1. SELF-TEST   

Attracting the main and auxiliary part together, press down “POWER" key to power 

them respectively and align the upper and lower laser points with the alignment bull’s 
eyes simultaneously. Wait for the completion of “SELF-TEST” and enter the 
measurement interface(100% displayed on the LCD). The following reasons for failed 
“SELF-TEST”(“Please refer to the manual” display on the LCD) : 