-9-
<
For CAN/CAN FD
>
Use DSUB9pin Monitor cable for CAN/CAN FD communications.
2.2 Connect to the Target Device
It is possible to select 2 measuring channels of separately from CAN FD/
CAN or CXPI. Connect Ch1 to the port of CAN1 or LIN1, and Ch2 to
the port of CAN2 or LIN2, using the cables for CAN or LIN attached to
the product.
■ Online Monitor
It is able to measure 2 channels simultaneously. Select CAN1 (CAN FD/
CAN) or LIN1, and CAN2 (CAN FD/CAN) or LIN2.
■ Simulation
Select 1 port from CAN1, CAN2, LIN1 and LIN2.
DSUB 9pin branch cable
<
Setting of Terminator
>
The jumper pin makes it possible to connect the terminator in the CAN
communications.
(
Factory setting: terminator OFF
)
JP1 (1-2 short): Connect the terminator (120 Ohm) to CAN1 port.
JP2 (1-2 short): Connect the terminator (120 Ohm) to CAN2 port.
JP1
JP2