LD30x

7.2.5 Speed from differential time between a Start and a Stop input

Input A operates as a start input and input B operates as a Stop input. The 
differential time between start and stop will be converted into the speed of the 
passing object.

Menu

Setting Range

Default

timE

Time

Enter a typical delay time which you expect between 
the start and stop signals.

000.001

999.999 sec

1.000 sec

diSPL

Display Value

Enter the speed you would like to see in the display 
when an object passes within above time.

000.001

999.999

1.000

dPoint

Decimal Point

Select the desired position as shown in the display.

000.000

no decimal point

000000

one decimal position

00000.0

---->
five decimal positions

0.00000

WAit

Wait Time

How long should the last result remain in the display 
before   it  returns  to  zero?  Set  the  desired   waiting 
time. With setting “0” the display will freeze and wait 
until to the next measuring cycle.

0.00

...

9.99

1.00

7.2.6 Linearisation points

The linearisation points will only appear with "RPM" or "Count" operation modes 
when Linearisation is enabled.

Menu

Setting Range

Default

P01_X

Linearisation point 1

X value of the first interpolation point.

-199999

... 999999

999999

P01_Y

Linearisation point 1

Y value of the first interpolation point.

-199999

... 999999

999999


P16_X

Linearisation point 16

X value of the 16. interpolation point.

-199999

... 999999

999999

P16_Y

Linearisation point 16

Y value of the 16. interpolation point.

-199999

... 999999

999999

For more details about linearisation please refer to section “8.1 Linearisation“ on 
page 30.

MAN LD30x E 1.0.odt

Set up procedure

20 of 40

Содержание LD30 Series

Страница 1: ... Model LD300 is a display only version model LD301 provides an analogue output in addition model LD302 further offers two presets and two switching outputs finally model LD303 is equipped with a RS 232 RS 485 serial interface for connection with a PC Table of contents 1 Safety summary 2 Identification 3 Mounting instructions 4 Electrical connections 5 Operating the front keys 6 Operator menu 7 Set...

Страница 2: ...ns in compliance with 2004 108 EC norm on electromagnetic compatibility following precautions must be taken before handling and installing the equipment discharge electrical charge from your body and tools which may come in touch with the device power supply must be stabilized without noise install EMC filters on device power supply if needed always use shielded cables twisted pair cables whenever...

Страница 3: ...istance For any information on the technical characteristics of the product refer to the technical catalogue 3 Mounting instructions WARNING Mount the unit with power supply disconnected Mount the display into the provided cut out approx 91 x 44 mm without panel clips Install panel clips on the display housing and screw until fixed 96 0 3 780 110 0 4 331 44 0 1 732 10 0 394 8 0 315 91 0 3 583 9 0 ...

Страница 4: ...ue output Special versions with TTL inputs option M8 provide a 5V aux output on terminal 7 instead of 24V Units with option P4 provide 24 42 VAC power supply instead of 115 230 VAC option PM LD302 Display unit with 2 presets and transistor outputs Special versions with TTL inputs option M8 provide a 5V aux output on terminal 7 instead of 24V Units with option P4 provide 24 42 VAC power supply inst...

Страница 5: ...rth connections may cause serious EMC problems 4 2 Aux voltage output Terminal 7 provides a 24 VDC 120 mA max auxiliary output for supply of sensors and encoders Units with TTL inputs option M8 provide a 5 VDC 120 mA auxiliary output on terminal 7 instead 4 3 Inputs A B and Reset In the basic set up menu see section 7 1 Basic settings on page 14 these inputs can be configured to PNP switch to or t...

Страница 6: ... V 10 V according to setting At the same time a current output 0 4 20 mA is available Both outputs refer to the GND potential and the signal polarity changes with the sign in the display The outputs provide a 14 bit resolution and the response time at each change of the measuring value is approx 7 msec fin 143 Hz The maximum current of the voltage output is 2 mA and the load on the current output ...

Страница 7: ... the board to the rear to remove the PCB from its housing ON DIP DIL Switch Removal of the back panel Location of the DIL switch RS 232 ON 8 9 10 RxD TxD GND RS 485 ON 8 9 10 A B GND WARNING 1 Never set DIL switch positions 1 and 2 or DIL switch positions 3 and 4 to ON at the same time 2 After setting the switch shift the print carefully back to its housing in order to avoid damaging the front pin...

Страница 8: ...e 14 Operational parameters set up see section 7 2 Operational parameters on page 16 Teach operation see section 5 3 Teach operation on page 10 5 1 Normal display state NOTE You can only change over to other operation states while the unit is in display state Change over to Key operation Basic set up Keep ENTER and SET down simultaneously for 3 seconds Operational parameter set up Keep ENTER down ...

Страница 9: ...ed blink This procedure allows setting all digits to the desired values After the most significant digit has been set the low order digit will blink again and you can make corrections if necessary With signed parameters the high order digit will scroll from 0 to 9 positive followed by and 1 negative 5 2 3 Saving settings To save the actual setting press the ENTER key which will also automatically ...

Страница 10: ...eaching of the interpolation points on page 32 5 4 Setting all registers to Default values At any time you can return all settings to the factory default values WARNING This action will reset all parameters to factory default values and your own settings will be lost You will have to repeat your individual set up procedure Factory default values are shown in the subsequent parameter tables see sec...

Страница 11: ... been switched on see on page 15 any key access first results as follows To access the menu you must press the following key sequence within 10 seconds otherwise the unit will automatically return to the normal display mode MAN LD30x E 1 0 odt Operating the front keys 11 of 40 ...

Страница 12: ... to be most suitable for simplification of the programming procedure The subsequent table shows the general structure of the menu Detailed descriptions of all parameters will follow in section 7 Set up procedure on page 14 6 1 Overview of basic parameters LD300 LD301 LD302 LD303 Type Application mode Type Application mode Type Application mode Type Application mode Input characteristics Input char...

Страница 13: ...t Wait Time Average Filter Average Filer Decimal Point Analogue Begin Analogue End LD302 Preselection 1 Preselection 2 Frequency Display Format Base Resolution Counter Mode Time Display Value Frequency Start Stop Scaling Factor Display Value Decimal Point Decimal Value Auto Reset Set Value Decimal Point Wait Time Wait Time Latch Function Reset Set Wait Time Average Filter Average Filer Decimal Poi...

Страница 14: ...NP NPN and the desired brightness of the LED display NOTE To access the Basic Set up press the ENTER and SET keys simultaneously for at least 3 seconds Menu Setting range Default tYPE Operation Mode rPm Tachometer frequency meter see section 7 2 1 RPM operation as tachometer or frequency counter rPm Baking time processing time indicator see section 7 2 2 Time display of baking or processing time r...

Страница 15: ...gs for preselections P_frEE LinEAr Linearisation mode For details please refer to sections 8 1 Linearisation and 8 2 Manual input or teaching of the interpolation points no The linearisation is switched off no Linearisation settings for the positive range only negative values will appear as a mirror 1 qUA Linearisation over the full numeric range 4 qUA with Tachometer mode and Counter mode only MA...

Страница 16: ...e unit will wait from one input pulse to the next before it sets the display to zero When you enter 0 the unit will wait forever and show the last result until it receives the next input 0 1 99 9 sec 1 0 The setting of this parameter automatically limits the minimum input frequency correspondingly With Wait set to e g 0 1 sec the unit will respond to frequencies 10 Hz only and all lower frequencie...

Страница 17: ...our display with above frequency at the input 1 999999 100 WAit Wait Time Define a waiting time this is the time in seconds that the unit will wait from one input pulse to the next before it sets the display to zero When you enter 0 the unit will wait forever and show the last result until it receives the next input 0 1 99 9 sec 5 0 The setting of this parameter automatically limits the minimum in...

Страница 18: ...ment St_SP Time count is active while input A is HIGH Hi_Loo Rising edge on input A starts count Rising edge on input B stops count St_SP Period time measurement Repeating display of the time between two rising edges on input A A_StSP rESEt Auto Reset no Time count goes on and never restarts No automatic Reset Use the Reset input to set to zero no Every start initializes a new count starting from ...

Страница 19: ...ered here 199 999 999 999 0 rESEt Reset Set Enable Fr u E No setting or resetting of the counter is possible no Set Reset by the front SET key front Set Reset by remote signal to the Reset input E_tErn Set Reset by front SET key and by external input Fr u E dPoint Decimal Point Select the desired position as shown in the display 000 000 no decimal point 000000 one decimal position 00000 0 five dec...

Страница 20: ...s 0 00000 WAit Wait Time How long should the last result remain in the display before it returns to zero Set the desired waiting time With setting 0 the display will freeze and wait until to the next measuring cycle 0 00 9 99 1 00 7 2 6 Linearisation points The linearisation points will only appear with RPM or Count operation modes when Linearisation is enabled Menu Setting Range Default P01_X Lin...

Страница 21: ... 999 9 999 0 000 GAin Analogue Gain Set the analogue stroke you desire setting 1000 means 10 Volts or 20 mA Setting 200 reduces the stroke to 2 Volts or 4 mA 00 00 99 99 10 00 See next page for more details Response time of the analogue output Operation mode Response time analogue output Tachometer frequency meter see section 7 2 1 RPM operation as tachometer or frequency counter on page 16 330ms ...

Страница 22: ...wo parameters any window of the whole display range can be mapped onto the analogue output The subsequent example shows how to convert the display range from 1400 to 2200 into an analogue signal of 2 10 volts 2 4 6 8 10 1000 2000 3000 4000 5000 Display Value Volts Analogue Output A ChAr 0 10 V AnAbEG 1400 OFFSEt 2 000 AnAEnd 2200 GAin 8 00 NOTE All settings refer to the scaled values which are sho...

Страница 23: ...e output and automatic setting to preset 1 when the display value reaches zero CHAr 2 Switching characteristics of output 2 __r GE __r GE See CHAr 1 __r LE See CHAr 1 _n_ GE See CHAr 1 _n_ LE See CHAr 1 __r 1 2 Output switches statically ON when display value Preset 1 Preset 2 _n_ 1 2 Output switches dynamically ON when display value Preset 1 Preset 2 HYSt 1 Hysteresis 1 adjustable hysteresis for ...

Страница 24: ...steresis GE Greater Equal Display Value Hysteresis effect with Greater Equal Preselection LE Lower Equal Hysteresis Display Value Hysteresis effect with Lower Equal It is possible to check up on the actual switching state of the outputs at any time For this just tap on the ENTER key shortly The display will then show for the next two seconds one of the following information Display Meaning 1_2oFF ...

Страница 25: ...ot contain a 0 because such numbers are reserved for collective addressing 0 99 11 S Forn Serial data format 7 E 1 The first character indicates the number of Data Bits The second character specifies the Parity Bit even or odd or none The third character indicates the number of Stop Bits 7 E 1 7 E 2 7 0 1 7 0 2 7 no 1 7 no 2 8 E 1 8 0 1 8 no 1 8 no 2 S bAUd Baud rate 9600 The following baud rates ...

Страница 26: ...r transmissions are shown in the table Baud Rate Minimum Cycle Time msec 600 384 1200 192 2400 96 4800 48 9600 24 19200 12 38400 6 S mod Serial Mode PC Operation according to communication profile see section 7 5 1 PC Mode on page 27 PC Transmission of string type 1 see section 7 5 2 Printer Mode on page 29 Print1 Transmission of string type 2 see section 7 5 2 Printer Mode on page 29 Print2 S Cod...

Страница 27: ...nd as shown on the right Leading zeros will be suppressed BCC represents a block check character generated from an Exclusive OR of all characters from C1 through ETX inclusively STX C1 C2 x x x x x x x ETX BCC STX Control character Hex 02 C1 Register code High Byte C2 Register code Low Byte x x x x x Register data ETX Control character Hex 03 BCC Block check character With incorrect request string...

Страница 28: ... reset a unit with unit No 11 Reset ON ASCII EOT AD1 AD2 STX C1 C2 Data ETX BCC Hex 04 31 31 02 36 30 31 03 34 Binary 0000 0100 0011 0001 0011 0001 0000 0010 0011 0110 0011 0000 0011 0001 0000 0011 0011 0100 Reset OFF ASCII EOT AD1 AD2 STX C1 C2 Data ETX BCC Hex 04 31 31 02 36 30 30 03 35 Binary 0000 0100 0011 0001 0011 0001 0000 0010 0011 0110 0011 0000 0011 0000 0000 0011 0011 0101 For more deta...

Страница 29: ...Data Carriage return X X X X X X CR The mode of activation of serial transmissions can be determined as follows Cyclic timed transmissions Set the S tim item see on page 25 to any value 0 010 sec Select the desired string type next to parameter S mod After exiting the menu the timed transmissions will start automatically Manual activation of transmissions Set the S tim item see on page 25 to 0 000...

Страница 30: ...y show in the display With parameters P01_y to P16_y you can specify which values you would like to display instead of the corresponding _x values This means e g that the unit will replace the previous P02_x value by the new P02_y value NOTE With respect to the consistency of the linearisation the x registers have to use continuously increasing values e g the x registers must conform to the constr...

Страница 31: ...like to display the clearance of the gate d but the existing encoder information is proportional to the angular information φ d φ 1 cos φ Incremental Encoder d d0 Display value P01_x P05_x P16_x P01_y P05_y P16_y P03_x P03_y P07_y P07_x Encoder Position φ Display d P15_y P15_x MAN LD30x E 1 0 odt Special functions 31 of 40 ...

Страница 32: ...it has switched over to the Teach mode To start the teach procedure please press again the CMD key within the next 10 seconds The display will then show P01_X With respect to the consistency required for linearisation all parameters from P01_x to P16_ y will be overwritten by suitable initial values first Initial values for P01_x and P01_y are 199999 and all other values will start with 999999 Pre...

Страница 33: ...f interpolation points Every interpolation point must be higher than its preceding point If this constraint is breached all 6 decimal points will blink automatically as a warning Pressing the CMD key will not store the illegal value but result in an error text E r r L O To exit the teach mode again you have the following two possibilities 1 Press the ENTER key for 2 seconds On the display you will...

Страница 34: ...rmat diSFor 0 3 0 1 0 09 Frequency Hz FrEqu 1 25000 100 5 0 10 Display Value diSPL 1 999999 100 6 0 11 Wait Time sec WAit 1 999 50 3 1 12 Average Filter FiLtEr 0 4 0 1 0 13 See section 7 2 3 Timer Stopwatch on page 18 Base bASE 0 6 0 1 0 14 Start Stop StArt 0 2 1 1 0 15 Auto Reset rESEt 0 1 0 1 0 16 Latch Function LAtcH 0 1 0 1 0 17 See section 7 2 4 Count Counter mode on page 19 Count Mode ModE 0...

Страница 35: ... B1 P7 x P07_H 199999 999999 999999 86 0 B2 P7 y P07_Y 199999 999999 999999 86 0 B3 P8 x P08_H 199999 999999 999999 86 0 B4 P8 y P08_Y 199999 999999 999999 86 0 B5 P9 x P09_H 199999 999999 999999 86 0 B6 P9 y P09_Y 199999 999999 999999 86 0 B7 P10 x P10_H 199999 999999 999999 86 0 B8 P10 y P10_Y 199999 999999 999999 86 0 B9 P11 x P11_H 199999 999999 999999 86 0 C0 P11 y P11_Y 199999 999999 999999 ...

Страница 36: ... Positions Characters Serial code Preselection Value 1 PrES1 199999 999999 10000 86 0 27 Preselection Value 2 PrES2 199999 999999 5000 86 0 28 Preselection Mode 1 CHAr1 0 5 0 1 0 29 Preselection Mode 2 CHAr2 0 5 0 1 0 30 Hysteresis 1 HYSt1 0 99999 0 5 0 31 Hysteresis 2 HYSt2 0 99999 0 5 0 32 9 5 Serial interface model LD303 See section 7 5 Model LD303 additional settings for the serial interface o...

Страница 37: ...This page intentionally left blank ...

Страница 38: ...This page intentionally left blank ...

Страница 39: ...This page intentionally left blank ...

Страница 40: ...ment release Description 1 0 1st issue Lika Electronic Via S Lorenzo 25 36010 Carrè VI Italy Tel 39 0445 806600 Fax 39 0445 806699 Italy eMail info lika it www lika it World eMail info lika biz www lika biz ...

Отзывы: