Operations Manual

Unlock the Power
of Alcohol Testing

Содержание FC10Plus

Страница 1: ...Lifeloc FC10Plus Operations Manual Unlock the Power of Alcohol Testing...

Страница 2: ...rride During an Automatic Test 10 Conducting a Manual Test 11 Conducting a Passive Test 11 Operation Breath Flow 12 Alcohol Curve 12 Test Results 12 Viewing Previous Test Results 12 Removing the Mouth...

Страница 3: ...Type 21 Wet Bath Calibration Wet Bath Simulator Set Up 22 Wet Calibration 22 23 Wet Calibration Check 23 24 Dry Gas Calibration Dry Gas Tank Set Up 24 Dry Gas Calibration 24 25 Dry Gas Calibration Che...

Страница 4: ...lus is a state of the art breath alcohol tester that is software based and incorpo rates unique cutting edge technologies Because of the advanced FC10Plus design Results on a positive test register in...

Страница 5: ...Function Button scrolls through menus and saves changes Execute Button enables you to perform the function Mouthpiece Removal Tab Large Graphic LCD Screen displays functions and results Power Button...

Страница 6: ...en container without using a mouthpiece On Board Memory Stores the last 10 tests Real Time Clock Stores time and date information with test results as well as calibration and cal check results User Se...

Страница 7: ...e case by pushing up on the battery door until it locks shut See page 27 for directions on using rechargeable batteries Press and hold the Power button on the bottom of the front of the unit until it...

Страница 8: ...taking a test If the subject just took a drink a 15 minute observation period in which they are not permitted to put anything in their mouth should be observed before testing This will ensure all res...

Страница 9: ...suf cient breath sample for the automatic test Passive Test is a quick screen to detect alcohol but is not designed to quantify the results Passive results are reported as POS if alcohol is detected...

Страница 10: ...ch 1 3 liters Read the result After taking a test the FC10Plus will display the results in large numbers on the display Press the Function button to return to the test mode Note This feature allows th...

Страница 11: ...de is pressure activated If the subject does not blow air into the mouthpiece the test can not be conducted Turn the FC10Plus on Press the Function button if necessary until the display reads PASSIVE...

Страница 12: ...he alcohol is graphed the test result is displayed The result will remain on the screen until the Function or Power button is pressed The last 10 test results are retained in memory Viewing Previous T...

Страница 13: ...Momentarily press Power to return to the testing mode Press the Function button until the display reads SETTINGS Press the Execute button Display reads TIME Press the Function button until the displa...

Страница 14: ...Automatic Manual and Passive tests print test results and check battery status and temperature Press the Function button until the display reads SETTINGS Press the Execute button Display reads TIME P...

Страница 15: ...it Press the Function button to gain access You now have access to all menus and settings Press the Function button until the display reads SETTINGS Press the Execute button Display reads SECURITY COD...

Страница 16: ...r to return to the testing mode Please note By selecting Test Order 3 you will disable passive testing User settings ow chart is shown on p 32 Test Order 1 Default Test Order 2 Test Order 3 Auto Test...

Страница 17: ...tton to save your setting Momentarily press Power to return to the testing mode Results Format Explained Results for Auto and Manual tests can be displayed in a NUMERIC or PASS WARN FAIL PWF format Pa...

Страница 18: ...thod You must rst set your FC10Plus to recognize which method or which Standard Type you will be using Once you choose the Standard Type the FC10Plus will store that information in memory and you do n...

Страница 19: ...the settings Momentarily press Power to return to the testing mode Once you set the standard you do not have to set it again unless you change solutions or elevation dry gas only Calibration Cal Check...

Страница 20: ...Function button until the display reads STANDARD TYPE Press the Execute button to choose between DRY GAS and WET BATH Press the Function button to save settings Momentarily press Power to return to t...

Страница 21: ...tions The time starts counting from your last calibration date not from the date you set it Choosing DISABLED turns this setting off Press the Function button to save your setting Momentarily press Po...

Страница 22: ...ulator manual for location Turn the FC10Plus on Attach a mouthpiece to the back of the unit Press the Function button until the display reads CALIBRATION then press the Execute button Display reads WE...

Страница 23: ...r securely to the output port on the simulator lid Refer to your simulator manual for location Turn the FC10Plus on Attach a mouthpiece to the back of the unit Press the Function button until the disp...

Страница 24: ...the regulator to the dry gas tank Attach the small tube to the output port on the regulator Securely t the mouthpiece adaptor to the small tube on the regulator Note Do not store the dry gas tank wit...

Страница 25: ...start Press the Execute button to proceed and start a 3 second countdown Press the Execute button again to take a sample Continue holding the Regulator button for another 3 seconds Stop pressing gas...

Страница 26: ...ressing gas ow will start Press the Execute button to proceed and start a 3 second countdown Press the Execute button again to take a sample Continue holding the Regulator button for another 3 seconds...

Страница 27: ...e the ability to use NiMH rechargeable batteries as well To use NiMH insert the batteries as directed and power on the unit Press the Function button until the display reads SETTINGS Press the Execute...

Страница 28: ...s been detected Move to a different location and try again Flow Error Retry Blow Steadily Exhalation not complete or interrupted Instruct subject to blow steadily as long as they can Invalid Calibrati...

Страница 29: ...to the Lifeloc factory Misuse abuse negligence or accidents Extended Service Plans are available for your FC10Plus These provide complete coverage for an additional year at a reasonable cost and incl...

Страница 30: ...d results or incorrect interpretation of results FC10Plus is not to be used by children under 12 years of age DO NOT DRINK AND DRIVE Lifeloc strongly recommends that no vehicle or machinery be operate...

Страница 31: ...t current listing of supplies for your FC10Plus go to www lifeloc com Click on Supplies FC Series or www lifeloc com store supplies fcseries html You can place your order online or call Customer Servi...

Страница 32: ...s Chart Settings Time Date Trigger Mode LCD Contrast Security Settings Test Order Results Format Fail Level Pass Level Exit Battery Type Display Settings Exit Execute Function Function Function Functi...

Страница 33: ...tion XXX Check XXX Calibration Standard Type Cal Standard Cal Time Type Cal Settings Check Time Exit Exit Lockout Type Execute Execute Execute Execute Function Function Function Function Function Func...

Страница 34: ...ity and ease of use as our priorities And our 5 Star Service is the fastest in the industry Lifeloc breath testers are used across the US and in over 35 countries We carry all of the accessories and s...