USER
MANUAL
20
51285-006
Home Use
T
C is optional for home use where it may be used for video recordings. There is no connection between
the PC and the T4 Amplifier unit, as all communication is accomplished wirelessly.
Mains Power
Medical Grade
Laptop Mains PSU
Patient Isolation Zone
Patient Event
Pushbutton*
AUX*
Sp0
2
*
Home use:
Wireless only
(no electrical connection to PC)
*Type BF
Laptop PC
optional for
home use
Trackit T4 Amplifier
with battery pack in bag
EEG Electrodes*
Patient
Figure 2: Connecting the T4 Amplifier - Home Use
3.2
Laptop installation and operation
The laptop must only be connected to the medical-grade laptop power supply supplied or
authorised by Lifelines. Do not use a standard laptop power supply.
Only use the laptop supplied or authorised by Lifelines.
1. Connect the power cord to the medical-grade power supply.
2. Connect the power supply output to the power input connector on the laptop.
3. Connect the power cord to mains power outlet.
NOTE:
The mains power cord serves as a power disconnect device. It should be installed near the equipment
and be easily accessible.
4. For laptop installation and operation refer to the manufacturer’s instructions supplied with it.
Do not touch simultaneously any accessible USB or other contacts on the PC and the patient.
If the USB cable is used in the home, the laptop and power supply must be placed 1.5m away from
the patient.
Connections and Usage
Содержание trackit T4