2.5
Use in the home environment
Where the equipment is intended to be used in the home, the unit should be operated in its bag where it is
protected against ingress of solid objects and water to a degree of IP22.
The laptop PC is optional in the home environment and may be used for video recordings. There is no cable
connection between the PC and the T4 Amplifier unit, as all communication is accomplished wirelessly.
Keep the equipment away from sources of heat.
Do not use mobile phones.
Do not allow pets or children to interfere with the equipment or sensor cables.
When the equipment is operated with or without its Bluetooth or WiFi on, other devices in the vicinity should be
moved away or turned off to reduce the likelihood of interference to the equipment or by the equipment.
The T4 may have internal radios fitted. These are approved industry-standard Bluetooth and WiFi types which
present minimal risk of reciprocal interference with other equipment.
2.6
Use with other equipment
Defibrillators and HF surgical equipment
The equipment is not defibrillator proof and should not be used in situations where a defibrillator is likely to be used.
The equipment should not be used with, or in the presence of, high frequency surgical equipment.
Other patient-connected equipment
When used simultaneously with other patient-connected equipment, for example a cardiac pacemaker or other
electrical stimulator, it is unlikely that a safety hazard will arise. However always consult the documentation
supplied with the other patient-connected equipment to ensure that all hazards, warnings and cautions are
considered before the equipment is used together.
Leakage current
This system is designed to comply with IEC 60601-1, the international standard for medical electronic equipment, which
specifies the permissible levels of leakage current. A potential hazard exists in the summation of leakage currents
caused by connecting several pieces of equipment together. Because this system can be used in conjunction with
standard electronic devices, the total leakage current should be tested whenever the system is modified.
There should be no electrical connections between the system equipment, which is powered via a medical grade
power supply, and any other equipment powered from another mains supply.
Installation and Maintenance