7�5 Setting Up an External Monitor
The Trackit M also can be used with an external monitor.
To set up an external monitor
1. Use the provided HDMI mini to HDMI cable to connect an external monitor to the HDMI port on the back of the
Trackit M.
HDMI Port
USB Port
Power Port
Fig 12: Trackit M Main Unit—Rear View
External equipment intended for connection to signal input, signal output or other connectors shall
comply with the relevant product standard e.g. IEC 60950-1 or IEC 62368-1 for IT equipment and
the IEC 60601-series for medical electrical equipment. In addition, all such combinations – Medical
Electrical Systems – shall comply with the safety requirements stated in the collateral standard IEC
60601-1- 1 or the general standard IEC 60601-1, edition 3, clause 16. Any equipment not complying with
the leakage current requirements in IEC 60601-1 shall be kept outside the patient environment i.e. at
least 1.5 m from the patient support or shall be supplied via a separation transformer to reduce the
leakage currents.
Any person who connects external equipment to signal input, signal output or other connectors has
formed a Medical Electrical System and is therefore responsible for the system to comply with the
requirements. If in doubt, contact qualified medical technician or your local representative.
An approved Separation Device (isolation device) is required when connecting an external monitor to
the Trackit M. The requirement for the Separation Device is defined in IEC 60601-1- 1 and in IEC 60601-
1, edition 3, clause 16.