Sma

User’s Manual v1.11

Содержание SmartBatch+

Страница 1: ...SmartBatch User s Manual v1 11...

Страница 2: ...er Interface 9 Graphing History 11 Operation 12 Clearing Mode 12 Labeling Mode 13 Switching Between Modes 14 Device Washing 14 Temperature Calibration 14 Replacing Electrodes 15 Replacing Fuse 15 Trou...

Страница 3: ...are loaded into the Sample Chamber where an electric field is formed between the Electrodes to drive charged molecules into the samples For each operation unique buffers are poured into the Sample Cha...

Страница 4: ...Power cord D 1 x Batch Staining Cup E 2 x Incubation Jar F 1 x Extended Drain Tube G 2 x Ring Stand H 2 x Sample Ring and 2 x Lock Ring I 2 x Extra Fuses J 2 x Mesh Bag Set K 2 x 12 bags for mouse br...

Страница 5: ...Ambient Conditions Operating Temperature Range 18 26 C Relative Humidity 80 Max Storage Temperature 18 26 C Storage Humidity 80 Max Altitude 2 000 m Electrical Data Rated Voltage Range 10 100 240 VAC...

Страница 6: ...ument to an electrical outlet according to local or country standard Do not switch on the instrument without having waited at least 10 seconds after switching it off Only use buffers sample cups and p...

Страница 7: ...0cm on each side to allow for proper airflow Image 3 Plug the power cord into the power entry module on the back of the device and then plug it into the appropriate wall outlet 4 Flip the power switch...

Страница 8: ...er towels from the chamber and dispose of them These are only here to soak up any remaining liquid during shipping 7 Pour 400 mL distilled water directly into the Chamber www lifecanvastech com info l...

Страница 9: ...ce is indeed rotating The software will also deactivate so that power cannot be cycled too quickly for some components 9 If you wish to use the optional Extended Drain Tube simply press the hard plast...

Страница 10: ...startup The bottom number shows the Set Temperature of the system The Set button will change the Set Temperature to the value that is currently displayed in the keypad If an invalid temperature setti...

Страница 11: ...d and then press the Set button If an invalid Timer value is input a sound will play and the value will go to the maximum allowable value This is the Keypad The number at the top is the current value...

Страница 12: ...s is the Faults Panel When the Case Lid is open the Cover Open indicator will light up When the lid is open the device will halt any Electrophoresis Power To turn this back on close the lid and press...

Страница 13: ...carefully with a gentle stream of water 9 Place the Clearing Cup into a beaker of distilled water for several minutes to finish washing the cup 10 Remove the Clearing Cup from the liquid and dump out...

Страница 14: ...lace the Batch Staining Cup into a beaker of distilled water to finish washing it 6 Remove the Batch Staining Cup from the liquid and dump out any water Carefully use a kimwipe to soak up any remainin...

Страница 15: ...pump for several minutes 5 Turn off Auxiliary Power and drain the liquid 6 Repeat steps 3 5 at least 3 more times or until the liquid in the Chamber is no longer bubbly Temperature Calibration We rec...

Страница 16: ...rode swap is safe 2 Drain out any liquid from the device 3 Pull the red or black wire out from the Electrode Assembly 4 Unscrew the 4 bolts that secure the Electrode Assembly to the Chamber 5 Wipe up...

Страница 17: ...en Case Lid is opened for safety Close Case Lid and press Reset Faults button to turn electrophoresis back on Overheat Measured tank temperature is above 50 C Auxiliary Power and Electrophoresis Power...

Страница 18: ...0 C voltage higher than 90V current higher than 1500mA Any damages due to fire earthquake rainstorm or other catastrophic events as well as damages arising from pollution or abnormal electrical supply...