Odyssey® Sa Imager
42) Press the eject button on the Odyssey Sa Imager to retract the plate carrier.
43) Turn on the BioStack instrument and wait for it to initialize.
BioStack Z-Axis Training
The final procedure is to adjust the vertical (Z-axis) travel of the BioStack
plate gripper. The goal is for microplates to be placed on the Odyssey Sa
Imager plate carrier from the optimal height. If the value of the vertical
travel distance on the Z-axis is too small, the plate will drop onto the
Odyssey Sa Imager plate carrier rather than being placed on it. If the
value of the vertical travel distance is too large, the plate will hit the
Odyssey Sa Imager plate carrier and the plate gripper arms will continue
to travel downward, which could result in the bottom of the gripper
assembly impacting the microplate as it rests on the Odyssey Sa Imager
plate carrier.
44) Start Image Studio Software on your computer. Open the application
menu, hover over
Instrument
,
Management
, and click
Admin-
istration
.
45) Click
BioTek BioStack
in the System Administration window.
Page 43