Section 7. Software Reference
Signal Strength Tab
The
Signal Strength
tab has a button labeled
Reset Signal Strength
that you can
click if you’ve decided your instrument optics are as clean as you can reasonably
make them, and you want to reset the signal strength to 100.
See
on page 8-7 for more information.
Coefficients Tab
The
Coefficients
tab displays factory-determined calibration coefficients, a factor for
correcting CO
2
measurements for band broadening due to the presence of water
vapor (A), and a zero drift correction factor (Z). The coefficient shown as XS (Cross
Sensitivity) compensates for slight cross sensitivity between CO
2
and H
2
O signals
7-22
LI-7200RS Enclosed CO₂/H₂O Analyzer
Содержание LI7200RS
Страница 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Страница 2: ......
Страница 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Страница 12: ...x...
Страница 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Страница 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Страница 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Страница 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 225: ......