![LI-COR LI7200RS Скачать руководство пользователя страница 132](http://html.mh-extra.com/html/li-cor/li7200rs/li7200rs_instruction-manual_1906669132.webp)
Section 6. Troubleshooting
Loose connection? Make sure the power cable connector is tight. Also check the
cable for damage and the connections to the power supply.
Blown fuse? Check the fuse, as described in
on page 5-8. Note
that a blown fuse usually indicates some other problem. If the fuse blows
repeatedly, carefully check the wiring.
Problems with the SmartFlux 2 System
System not visible on your network (cannot connect)
l
Check the Ethernet Activity LED light on the SmartFlux 2 module. If it is blink-
ing rapidly, this indicates the system is communicating over the network.
l
Check the Ethernet cable and connections to the SmartFlux 2 module and in
the rest of your EC system.
l
Check the power supply connections and correct any faulty connections.
l
Reboot the SmartFlux 2 System by pressing the reset button.
l
Reboot the LI-7550 Analyzer Interface Unit.
System not connected to the sonic anemometer
l
If you have renamed the SmartFlux 2 System, does it have a valid name? The
SmartFlux 2 System name can only include upper and lower case letters (A to
Z; a to z), numbers (0 to 9), the dash character (-), and the period (.). All other
characters are invalid and will prevent the system from working as expected.
One indicator of this is that the sonic anemometer status is displayed as dis-
connected in the gas analyzer software main window.
l
Invalid names: Flux_Site_01 (invalid underscore _ characters); Micro-
met@TheSwamp (invalid @ character).
l
Valid names: smart03-00901 (the default name).
l
For manually configured Anemometers, check to see that the Sonic Anem-
ometer is set to the proper baud rate.
l
Check the return signal LED lights next to the sonic anemometer data port. If
they are flashing or solid, this indicates that the SmartFlux 2 System is com-
municating with the sonic anemometer.
l
Check the connections on the sonic data input plug wiring and tighten any
loose fittings.
6-2
LI-7200RS Enclosed CO₂/H₂O Analyzer
Содержание LI7200RS
Страница 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Страница 2: ......
Страница 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Страница 12: ...x...
Страница 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Страница 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Страница 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Страница 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 225: ......