Section 5.
Maintenance
The section describes maintenance procedures. A good maintenance plan will
ensure good performance of the instrument, reduce data gaps, and give you greater
confidence that the measurements truthfully represent physical processes. FluxSuite
Software can help immensely with your maintenance plan.
1
Eddy Covariance Site Maintenance Schedule
When you first deploy the instrument:
Check the flow drive percent and record it. This will help you determine if the
intake filter needs to be cleaned.
Check the signal strength and record this for a baseline. This will help you determ-
ine when the optics should be cleaned.
Every day or every few days:
Check the overall performance of the instruments, including the measured values
and diagnostic information. This will ensure that you don't lose data (or that you
lose less data) if something is wrong.
Check the measured values. Air temperature, pressure, sonic temperature, dew
point, gas concentrations, covariances and fluxes. Any unexpected readings may
indicate an issue.
Check the diagnostics. Signal strength, detector temperature, chopper housing tem-
perature, and thermocouples.
1
Please forgive this shameless plug. But do check out www.fluxsuite.com.
5-1
LI-7200RS Enclosed CO₂/H₂O Analyzer
Содержание LI7200RS
Страница 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Страница 2: ......
Страница 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Страница 12: ...x...
Страница 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Страница 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Страница 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Страница 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 225: ......