Replacing the Source and Detector in
the LI-8x0 Analyzers and LI-8100/A

This document describes how to replace the optical source
and detector on the LI-830, LI-850, and LI-870 gas analyzers,
legacy LI-800, LI-820, LI-840, and LI-840A gas analyzers, and
the LI-8100/A Automated Soil CO

2

Flux System.

Part Numbers

Source

Detector

14 cm
Bench

5 cm
Bench

LI-850 / LI-
870

800-902

800-918

800-904

n/a

LI-830

800-902

800-916

800-904

n/a

LI-840/A

800-902

800-906

800-904

n/a

LI-820

800-902

800-903

800-904

800-905

LI-800

800-902

800-903

800-904

800-905

LI-8100/A

8100-902 800-906

800-904

n/a

Contents

LI-800, LI-820, LI-830, LI-840/A, LI-850, and LI-870

1

LI-8100/A Source and Detector Replacement

3

LI-800, LI-820, LI-830, LI-840/A, LI-850,
and LI-870

The optical source and detector may need to be replaced if
one fails. The life expectancy of the source is typically
>20,000 hours, or >2 years of continuous operation, but this
will vary depending upon the number or warm-up cycles,
vibration, and other factors. Typically, detectors need to be
replaced rarely, if ever.

Disassembly

You'll need #1 or #2 Phillips screwdriver and a clippers or
utility knife. The installation, including calibration, may take
an experienced technician 30 minutes to 1 hour. If the instru-
ment has been running recently, the bench may be hot to the
touch. Let it cool for 5 or 10 minutes after powering off if
needed.

Note:

Be sure that you are properly grounded to avoid elec-

trostatic discharge that can damage the electronics. Use an
anti-static wrist strap, electrostatic discharge grounding mat,

or occasionally touch bare metal that has a clear path to
ground, such as an unpainted computer case.

1

Power off the gas analyzer and remove the top cover
(LI-800, LI-820, LI-830, LI-840/A and LI-850) or open
the case (LI-870 only).

The LI-800, LI-820, LI-830, LI-840/A and LI-850 top cover
is held in place by 6 screws. Loosen the screws and lift off
the cover. If there is a ribbon cable connecting the top
panel, be careful not to strain it. The LI-870 is opened by
removing two screws that hold the case closed; then
remove four screws that secure the optical bench cover.

2

Unplug the source and detector ribbon cables from
the source and detector.

3

Lift the optical bench out of the foam insulation.

Replacing the Optical Source

With the optical bench free of the case (except the tubing),
you can remove the source from the optical bench.

1

Remove the tube from the source.

The simplest way is to cut the tubing as close to the hose
barb as possible.

2

Remove the four larger screws from the optical
source.

1

Installation Guide

Содержание LI-8 0 Series

Страница 1: ...experienced technician 30 minutes to 1 hour If the instru ment has been running recently the bench may be hot to the touch Let it cool for 5 or 10 minutes after powering off if needed Note Be sure that you are properly grounded to avoid elec trostatic discharge that can damage the electronics Use an anti static wrist strap electrostatic discharge grounding mat or occasionally touch bare metal that...

Страница 2: ... Make sure that the foam insulation on the inside top cover is positioned over the optical bench it is required for thermal stability Reconnect the top cover ribbon cable and then install the top cover 10 In the LI 830 LI 850 and LI 870 reset the source hours to 0 The LI 830 LI 850 and LI 870 record the number or oper ating hours for the optical source You should reset this count after replacing t...

Страница 3: ...ations you will measure A dual span however will bring the instrument performance within specifications across the full measurement range If you do not have zero and span gases send the instrument back to LI COR for a factory recalibration LI 8100 A Source and Detector Replacement The LI 8100A Soil Gas Flux System the source and detector on this optical bench can be replaced if they fail These ins...

Страница 4: ...of the white panel 7 Lift the assembly out of the yellow case and turn it upside down 8 Remove the 6 screws from the optical bench cover 9 Leaving the hoses connected to the optical bench lift the bench out of the foam casing 10 There are four screws on the source and detector housings that must be removed Remove only the screws from the assembly you are repla cing Remove the four screws in the co...

Страница 5: ... of the bench Note that the orientation of the optical bench cyl inder is not important either end can be inserted into the source or detector housing 15 Replace the tubing Any tubing that you removed should be replaced with a piece of equal length which can be cut from the tubing that is included in the instrument spares kit 16 Re assemble the LI 8100 A case Make sure that the foam insulation on ...

Страница 6: ...6 ...

Страница 7: ...7 ...

Страница 8: ...6 U S and Canada envsales licor com LI COR Distributor Network www licor com envdistributors Regional Offices LI COR Biosciences GmbH Siemensstraße 25A 61352 Bad Homburg Germany Phone 49 0 6172 17 17 771 envsales gmbh licor com LI COR Biosciences UK Ltd St John s Innovation Centre Cowley Road Cambridge CB4 0WS United Kingdom Phone 44 0 1223 422102 envsales UK licor com ...

Отзывы: