Section 2. Getting started
Instrument display
The instrument display panel provides useful information at a glance.
l
Serial Number and IP Address:
The instrument serial number (hostname) and
network-assigned IP address (IPv4) rotate for 5 seconds each. The serial number
is also printed on the case label. If the network cable is not plugged in, the dis-
play will give the IP address for the Wi-Fi network (if enabled).
l
Network Status:
When visible (
), the instrument is connected to a network.
The network cable is installed between the instrument network port and a port
on a local area network.
l
Wi-Fi Status:
When Wi-Fi is enabled (indicated by the
symbol), the instru-
ment broadcasts a Wi-Fi network with the same name as the instrument serial
number (and hostname).
l
Battery Status:
Indicates the charge status ( ). A solid black battery indicates a
full charge; an outline indicates a discharged battery. The battery icon is not dis-
played when no batteries are installed.
l
Power Status:
Indicates the power source (AC if power cable connected).
l
Performance Status:
Indicates performance status of the instrument.
l
Gases:
The gases measured and their concentrations.
2-8
LI-7820 N₂O/H₂O Gas Analyzer
Содержание LI-7820
Страница 1: ...LI 7820 N2O H2O Trace Gas Analyzer Instruction Manual ...
Страница 2: ......
Страница 10: ...viii ...
Страница 28: ...Section 2 Getting started 2 12 LI 7820 N O H O Gas Analyzer ...
Страница 36: ...Section 3 Applications of the LI 7820 3 8 LI 7820 N O H O Gas Analyzer ...
Страница 52: ...Section 4 Software overview 4 16 LI 7820 N O H O Gas Analyzer ...
Страница 66: ...Section 6 Maintenance 6 8 LI 7820 N O H O Gas Analyzer ...
Страница 84: ...Appendix B Additional safety information B 6 LI 7820 N O H O Gas Analyzer ...
Страница 91: ......