Section 6. Maintenance
Replacing the air inlet filter
The air intake has filters that may become contaminated periodically. We recom-
mend replacing filters every year, or more often if you are sampling dirty or dusty
air. The filters also should help prevent water from being drawn into the gas ana-
lyzer.
1
Power off the instrument.
2
Remove each of the six screws that secure the air intake and outlet.
They are installed with thread locking compound, which will make them difficult to
remove.
3
Remove the O-rings, the filter support disks, and the filter.
Use a toothpick to work them out.
4
Inspect the O-rings and replace them if they are damaged or compressed.
5
Install the replacement filters and reassemble the system.
Install
two
filters (part number 9978-234) between the support disks. The two metal
discs support the filter(s) — one on each side. Align the centers of the filters and the
metal disks. Assembly is the reverse of disassembly. You do not need to use thread
locking compound on the threads. Tighten them snugly.
6-4
LI-7815 CO₂/H₂O Gas Analyzer
Содержание LI-7815
Страница 1: ...LI 7815 CO2 H2O Trace Gas Analyzer Instruction Manual...
Страница 2: ......
Страница 10: ...viii...
Страница 30: ...Section 2 Getting started 2 14 LI 7815 CO H O Gas Analyzer...
Страница 38: ...Section 3 Applications of the LI 7815 3 8 LI 7815 CO H O Gas Analyzer...
Страница 42: ...Section 4 Software overview 4 4 LI 7815 CO H O Gas Analyzer...
Страница 54: ...Section 4 Software overview 4 16 LI 7815 CO H O Gas Analyzer...
Страница 62: ...Section 5 Troubleshooting 5 8 LI 7815 CO H O Gas Analyzer...
Страница 70: ...Section 6 Maintenance 6 8 LI 7815 CO H O Gas Analyzer...
Страница 88: ...Appendix B Additional safety information B 6 LI 7815 CO H O Gas Analyzer...
Страница 92: ...Appendix C Specifications C 4 LI 7815 CO H O Gas Analyzer...
Страница 97: ......