Troubleshooting
IRGA Problems
20-14
Using the LI-6400
20
If the flow did increase substantially, the problem is either with the soda lime
tube, or else in the line between the air intake and the soda lime tube. To test
the soda lime tube itself, remove the desiccant tube (that will make the flow
be highest) and put the soda lime tube in the desiccant tube location. If the
flow doesnÕt change much between having nothing in the desiccant tube lo-
cation, and having the soda lime in the desiccant location, that means the soda
lime tube is ok, and the problem is in the internal tube connecting the air inlet
and the soda lime location. Open the console, and inspect that line; it may
have debris in it.
If the problem is the soda lime tube, see
3
Remove the desiccant tube
If this improves the flow, the problem is in the desiccant tube. To find a block-
age within the desiccant tube, see
If this doesnÕt fix the problem, open the console. (See Figure 20-10 on
page 20-38.) There might be some sort of blockage between the air filter
(which is next after the desiccant tube) and the pump. Pull the air intake line
from the pump, and see if that increases the flow rate. If it does, youÕve brack-
eted the problem.
The problem could also be a bad pump or diaphragm, but this has been quite
rare. See
Pump Status: ERR
This can only happen with the 6400-01 installed, and then only when some-
thing is restricting the pump. Check for a blockage in the flow path on the
vacuum side of the pump, as described above.
IRGA Problems
ÒIRGAs Not ReadyÓ Message
The ÒIRGAs Not ReadyÓ message will show up in New Measurements mode,
and also during the IRGA zeroing and span setting routines, whenever one or
more of the gas analyzers is indicating a problem (see discussion under
on page 14-12). What triggers this message is Òtoo
muchÓ light attenuation in the non-absorbing reference wave bands for CO
2
and H
2
O. Causes:
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...