Calibration Issues
CO2 and H2O Analyzers
18-12
Using the LI-6400
18
Adjusting CO2_R (use
-¯
)
Mch:ON
CO2Rµml
457.2 449.0 -99.9 -99.9
CO2_R CO2_S H2O_R H2O_S Done
Fan:sF
1.009 1.007 .997 .981
CO2Sµml TD_R_¡C Td_S_¡C
Active IRGA. To change which
IRGA you are adjusting, use
the FCT keys, or else use
®
and
¬
.
Press the function key that
corresponds to the IRGA
you wish to adjust.
Span factors for each of the
IRGAs. Press
-
or
¯
to adjust
the reading (and this value) up
or down for the active IRGA.
These values are
G
cr
,
G
cs
,
G
wr
,
and
G
ws
used in Equations
(14-5) through (14-9) starting
on page 14-5.
Press
F
or
S
to adjust the fan
speed (slow and fast) and
M
to toggle the match valve.
Figure 18-6. Adjusting the reference CO
2
span. The span factor (should always be near 1.0) is used as a mul-
tiplier of the IRGA voltage. Pressing
-¯
increases or decreases the span factor for the active IRGA by a factor
of 0.001. To change the span factor by 0.010, hold
shift
down while you press
-
or
¯
. To change the active
IRGA, press the
F1
through
F4
, or else use
¬
or
®
.
LI-6400
PSH-0001
PORTABLE
PHOTOSYNTHESIS SYSTEM
MODEL
SR. NO.
LI-6400
PSH-0001
PORTABLE
PHOTOSYNTHESIS SYSTEM
MODEL
SR. NO.
If you connect only to the sample port,
then Match must be ON for the air to
get to the reference analyzer as well.
If you split the flow, Match can be Off.
(By turning Match ON and Off while
watching the reference concentration,
you can see the influence of the sam-
ple cell and chamber on the air stream)
A
B
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...