Making Measurements
Operational Hints
4-50
Using the LI-6400
4
¥
Match the IRGAs
As the differentials become small, the effect of any offset error is magnified.
Using Closed Mode
We have experimented with using a closed system technique in the LI-6400
to handle low rates. This is accomplished by turning off the pump for 10 or
15 seconds, and measuring the rate of change of CO
2
with time, and H
2
O with
time. The LI-6400 can be programmed to accommodate closed measurements
interspersed at will among normal, open measurements. However, our tests
indicate that the repeatability and accuracy of the closed technique is no better
(and sometimes worse) than using the open method with a low (100 µmol s
-1
)
flow rate. Contact LI-COR for further information.
Humidifying Incoming Air
The LI-6400Õs humidity control balances the leafÕs transpiration with drier in-
coming air to maintain a desired humidity (see
7-7). How dry the incoming air is depends on the manual adjustment knob of
the desiccant tube. The limitations of this approach become apparent when
measuring a small leaf area and/or low transpiration rate when it is desired to
have high humidity in the chamber. Another source of water besides the leaf
is needed.
One solution to this problem if you are using the 6400-01 CO
2
Mixer is to add
a small amount of water (10 ml) to the soda lime tube (Figure 4-14). After
about an hour of subsequent use, the water output becomes quite stable, and
remains so for many hours thereafter. The soda lime used for the experiment
in Figure 4-14 is supplied by W.R. Grace & Co, 5225 Phillip Lee SW, Atlan-
ta, GA 30336 (and is available from LI-COR as part #9960-071).
Caution
When adding water to soda lime, do it slowly, letting the chemical absorb the
liquid, and then let the tube sit for a few minutes before pumping air through
it. The goal is to avoid having any liquid get out of the tube, which can cause
oxidation of downstream metal components. Do not add any more than 10 ml.
of water to the tube.
Another solution is to use a brand of soda lime that has a higher water output.
A good example is that supplied by BDH Laboratory Supplies, Poole, BH15
1TD, England (product number 33115 AX). In situations where you need
fairly dry input air to the chamber, however, this material should be avoided,
as it can saturate the desiccant tube in an hour or two, making frequent chang-
es necessary.
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...