LI-600 Porometer/Fluorometer Instruction Manual
9-6
where
V
is the voltage read from the sensor,
Z
is a user-
settable zero parameter and
S
is a user-settable span para-
meter and
f(x)
is a 3rd or 4th order polynomial.
RH sensors
The RH Sensors have a small sensitivity to temperature
that must be accounted for. A temperature correction
occurs both in the zero setting and in the calibration
function. The sample RH sensor also must be corrected
for a match offset, as shown in
on page 9-8. We will use the terminology of
RH
r
and
RH
s
to indicate sensors before the cuvette (ref-
erence) and sensors after the cuvette (sample). Note that
a match offset is applied to the
RH
s
value (equation
9-12a
9-12b
where
m
x
and
b
xz
are factory-determined coefficients and
T
ref
is a thermistor embedded in the block adjacent to
the RH sensors.
Converting RH sensor voltage to RH (%) requires 4 fact-
ory-determined coefficients and the zero and span set-
ting for each sensor.
9-13a
9-13b
where subscript
r
and
s
represent the reference and
sample RH sensors, respectively.
is the sensor
voltage,
is the user-settable sensor zero,
is the
user-settable span (nominally 1.0).