LI-600 Porometer/Fluorometer Instruction Manual
9-14
9-40
where
P
freq
is the current pulse frequency (2 to 750,000
Hz),
P
width
is the pulse width (667 ns) and
Q
peakx
is the
peak intensity of the left and right LEDs, set by a factory
calibration.
The peak intensity is ~10,000 µmol m
-2
s
-1
and is cal-
culated from a calibration coefficient set at the factory.
Frequencies for steady-state measurements will be very
low (4 to 8 Hz), while frequencies during the flash will
be much higher (250 to 750 kHz).
Implementation on the LI-600
TheLI-600 calculates parameters as described.
Calculating light intensity from LED calibration
coefficients
Each LED has two calibration coefficients: a DAC set
point (
DAC
x
) and a parameter relating DAC setting to
light intensity (
Q
calx
). The peak intensity of each LED is
calculated from these coefficients:
9-41a
9-41b
A given configuration provides a steady-state mod-
ulation frequency
Mod
rate
in Hz and a desired flash
intensity for rectangular and phases 1 and 3 of multi-
phase flashes
flash_intensity
in µmol m
-2
s
-1
. The instru-
ment will determine the flash modulation frequency
based on desired user intensity and calculated peak
intensity (assuming both LEDs have the same intensity)
from: