For the LI-7500A/RS/DS or LI-7200/RS and file transfer utility
This procedure configures ports for the LI-7500A/RS or LI-7200/RS (with the
SmartFlux 2 System), or the SmartFlux 3 System (with the LI-7500DS). It also
opens a port for the file transfer utility, which makes it possible to copy data files
from the device. In ACEmanager™, under
Security > Port Forwarding
, click
Add
More
and enter the information.
Add port 7200 for the gas analyzerfile transfer utility are 7200 and 22.
If using the SmartFlux 2 System (not SmartFlux 3), open an additional port to
allow files to be copied from the LI-7550 USB drive.
Public Start Port
Public End Port
Protocol
Host IP
Private Start Port
7200
0
TCP
192.168.13.201
7200
7222
0
TCP
192.168.13.201
22
5222
1
0
TCP
192.168.13.200
22
Table 4-1
. Ports for the LI-7500A/RS or LI-7200/RS and SmartFlux 2 System, the
LI-7500DS and SmartFlux 2 System, and the file transfer utility used by both.
1
Used only to make it possible to copy files from the SmartFlux 2 System USB drive. Not required for the
SmartFlux 3 System.
Section 4. Configuring the RV50X
4-7
For the LI-7500A/RS/DS or LI-7200/RS and file transfer utility
Содержание AirLink RV50X
Страница 1: ...Using the AirLink RV50X with LI COR Eddy Covariance and Soil Gas Flux Systems Instruction Manual ...
Страница 2: ......
Страница 6: ...iv ...
Страница 10: ...viii ...
Страница 14: ...Section 1 Introduction to the communication system 1 4 AirLink Cellular Communication ...
Страница 28: ...Section 2 Initial assembly 2 10 AirLink Cellular Communication ...
Страница 36: ...Section 3 Configuring instrument IP addresses 3 8 AirLink Cellular Communication ...
Страница 52: ...Section 4 Configuring the RV50X 4 16 AirLink Cellular Communication ...
Страница 66: ...Section 6 Troubleshooting 6 4 AirLink Cellular Communication ...
Страница 68: ...Section 7 Maintenance 7 2 AirLink Cellular Communication ...
Страница 89: ......