7200-101 Blower Motor Replacement
This installation guide describes how to replace the 7200-
101 Flow Module blower motor (part number 9972-039).
The procedure takes about 30 to 60 minutes.
Tools Required:
Hex keys: 1/16", 7/64", and 5/32" Wrenches: 3/16" and 1/4"
Disassembly
1
Power off the flow module and disconnect the power
cable.
2
Place the flow module on a level work surface.
Remove the
adjust
knob using a 1/16" hex key.
3
Remove the four cap screws. Save the screws for re-
use.
4
Lift the panel straight up. Disconnect the power plug
and set the panel aside.
5
Use the 1/4" wrench to remove the three screws that
secure the blower plate. Separate the blower plate
from the blower assembly. Save the screws for re-
use.
1