5. 

Maintenance and Service 

5.1 

Maintenance 

The GRAPHIX controller does not require any special maintenance work. 
 

5.2 

Configuration 

With the help of this parameter group they have the possibility of securing and of restoring 
your  system  parameters.  In  addition  a  suitable  memory  at  the  USB  interface  must  be. 
Further resetting of the system parameters is possible on factory-installed settings. 
 

5.2.1 

Save Data 

To save the configuration parameters of your GRAPHIX controller proceed as follows: 

 

  Connect the USB stick with the USB socket at the front side of the equipment. 

  In the measurement mode, tap the touchscreen for approximately 1 second. 

o

  The main menu with an overview of the parameter groups is displayed.  

  To scroll, use the button 

o

  On the next page, main menu page 2/2 is displayed. The current page number is 

displayed in the upper section of the display. 

  In the main menu 2/2, tap on the parameter group Configuration. 

  In the parameter group Configuration, tap on the button 

Save Data

o

  The rest process starts. 

o

  On the memory at the USB interface the folder RECOVERY is installed. 

o

  Configuration data are stored in a file with the format rescue.txt on the USB stick in 

the folder RECOVERY. 

o

  After the reset process has been completed the instrument will automatically restart 

with the default parameters. 

o

  The GRAPHIX controller will now be ready for operation again. 

  When not wanting to run the reset, tap the button 

 to cancel the reset process. 

o

  The main menu is then displayed again. 

 

5.2.2 

Restore Data 

To  restore  the  saved  configuration  parameters  secured  on  an  USB  stick  in  the  listing 
RECOVERY in the file rescue.txt proceed as follows: 

 

  Connect the USB stick with the USB socket at the front side of the equipment. 

  In the measurement mode, tap the touchscreen for approximately 1 second. 

o

  The main menu with an overview of the parameter groups is displayed.  

  To scroll, use the button 

o

  On the next page, main menu page 2/2 is displayed. The current page number is 

displayed in the upper section of the display. 

  In the main menu 2/2, tap on the parameter group Configuration. 

  In the parameter group Configuration, tap on the button 

Restore Data

o

 

The question „Are you sure?” is displayed to ensure that you really want to run the 
reset. 

  When not wanting to run the reset, tap the button 

 to cancel the reset process. 

o

  The main menu is then displayed again. 

  To run the process, tap the button 

Restore now

o

  The rest process starts. 

o

  After the reset process has been completed the instrument will automatically restart 

with the default parameters. 

 

Содержание 230680V01

Страница 1: ...Catalog No 230680V01 230681V01 230682V01 GRAPHIXONE GRAPHIXTWO GRAPHIXTHREE Vacuum Gauge Controller Operation Manual GA300553102_002_C2...

Страница 2: ......

Страница 3: ...Symbols 10 4 4 1 Symbols for the Controls 10 4 4 2 Symbols for the Language Selection 11 4 4 3 Status and Warning Symbols 11 4 5 Measuring Mode 12 5 Maintenance and Service 13 5 1 Maintenance 13 5 2 C...

Страница 4: ...HIX controller is a display and operating unit with a graphic user interface for sensors with an analogue or digital interface of the company Leybold GmbH or from other manufacturers Depending on the...

Страница 5: ...sume any responsibility or warranty in case the operator or third persons do not observe the information given in this document do not use the product as intended modify the product in any way convers...

Страница 6: ...h electricity NOTICE General information pointing to further information respectively reference sections 2 3 Basic Safety Information During all work like installation maintenance and repair activitie...

Страница 7: ...measurement channel 2 for GRAPHIX TWO and THREE only for sensors CH2 A and CH2 B E1 E2 Connection of measurement channel 3 for GRAPHIX THREE only for sensors CH3 A and CH3 B F Relay output connection...

Страница 8: ...sible You may select between the following display modes Normal Default display mode which provides all important information on the connected sensors Chart Graphic display of the pressure history of...

Страница 9: ...e functions In this way you may configure your GRAPHIX controller and also utilise further functions offered by the instrument Channel Menu Buttons Briefly touching the desired channel invokes the cha...

Страница 10: ...Enabling display mode Leak Test Speedo Channel 1 Enabling display mode Speedo for channel 1 Speedo Channel 2 Enabling display mode Speedo for channel 2 Speedo Channel 3 Enabling display mode Speedo f...

Страница 11: ...ge RU Russian Polish Select menu language PL Polish Turkish Select menu language TR Turkish Table 4 Symbols for language selection 4 4 3 Status and Warning Symbols Symbol Designation Explanation Statu...

Страница 12: ...in specified measuring range Measuring value and additional warning symbol in the display mode Normal and Speedo or Measuring value displayed by a yellow font in the display modes Chart Big and Leak...

Страница 13: ...n the USB stick in the folder RECOVERY o After the reset process has been completed the instrument will automatically restart with the default parameters o The GRAPHIX controller will now be ready for...

Страница 14: ...Setup o The question Are you sure is displayed to ensure that you really want to run the reset When not wanting to run the reset tap the button to cancel the reset process o The main menu is then disp...

Страница 15: ...yed again To run the update tap the button Start update o On the memory at the USB interface is installed the folder RECOVERY o Configuration data are stored in a file with the format rescue txt on th...

Страница 16: ...measurement channel Acknowledge this message by selecting and terminating the channel menu at the corresponding channel Analog sensor signal out of range too high Error H 1 4 The measurement signal f...

Страница 17: ...memory stick Error usb file closing none 3 File on USB stick cannot be stored Check the attached USB memory stick Error usb write none 4 USB stick not recordably Check the attached USB memory stick Er...

Страница 18: ...n The setup value is saved To exit tap the button o The parameter selection display is displayed again 6 3 Help in Case of Malfunctions If the malfunction persists even after having exchanged the sens...

Страница 19: ...ry room During storage the following ambient conditions need to be maintained Ambient temperature 20 60 C Humidity of the air As low as possible Preferably in a sealed plastic bag with desiccant 7 3 W...

Страница 20: ...rvice zh leybold com Spain Leybold Spain S A C Huelva 7 E 08940 Cornell de Llobregat Barcelona Sales T 34 93 666 43 11 F 34 93 666 43 70 sales ba leybold com Service T 34 93 666 46 11 F 34 93 685 43 7...

Отзывы: