5.2 Analog Output
The transmitter can provide pressure measurement output as an analog voltage value or RS232 digital value. For
details about the digital interface please refer to the RS232 Instruction Manual 300544665. The analog output is
linear with pressure as shown in the graph below.
P
(
Torr
)
=
Pressure
in
Volts
*
Device Full Scale
Range
in
Torr
10
P
(
mbar
)
=
1
.
3332
*
Pressure
in
Volts
*
Device Full Scale
Range
in
Torr
10
P
(
Pascal
)
=
133
.
32
*
Pressure
in
Volts
*
Device
Full Scale
in
Torr
10
Pressure Units Conversion
Transmitter Full-Scale
Range - Torr
mbar
Pascal
0.1
0.133322
13.33224
1
1.333224
133.3224
10
13.33224
1333.224
20
26.66447
2666.447
100
133.3224
13332.24
1000
1333.224
133322.4
0.00%
20.00%
40.00%
60.00%
80.00%
100.00%
0.00
2.00
4.00
6.00
8.00
10.00
Pe
rcen
ta
ge
of
Fu
ll
Sc
al
e
Output Voltage - VDC
CERAVAC Output vs. % Full Scale