8.3.7
SAC (Save Actual Configuration)
Saving current configuration for switching threshold, sensor and general parameters.
S:
SAC<CR>
E:
OK<CR>
8.3.8
RSA (Read Serial Adress)
Read adress for RS485.
S:
RSA<CR>
E:
a<CR>
Parameter
Description
a
Adress 1
– 126 (01 – 7E)
(output in the HEX format)
8.3.9
SSA (Set Serial Adress)
Setting adress for RS485.
S:
SSA[a]<CR>
E:
OK<CR>
Parameter
Description
a
Adress 1
– 126 (01 – 7E)
(entry must be in the HEX format)
8.3.10
RGP (Read General Parameter)
Reading instrument settings.
S:
RGP<CR>
E:
a[,][TAB]b[,][TAB]c[,][TAB]d[,][TAB]e[,][TAB]f[,][TAB]g <CR>
Parameter
Description
a
Unit
0 = mbar
1 = Pa
2 = Torr
b
Analog output
0 = CM 31 mode
1 = CM 51 mode
c
Number of displayed digits
0 = 2
1 = 3