42
GA130001441_002_C0
– 09/2016
8.3
Command Set (Mnemonics)
8.3.1
Command Overview
Read Commands
Description
RPV
Read Pressure Value.
Read the pressure value for a measurement channel.
RVN
Read Version Number
Read the instrument software version number.
RSS
Read Setpoint Status
Read the status of the switching thresholds for a channel.
Table 21
– Mnemonics for read commands
Write Commands
Description
SHV
Set HV on/off
Switch the high voltage on or off for a channel.
SKL
Set Key Lock on/off
Switch the keylock on or off.
SAC
Save Actual Configuration
Save the current configuration for switching threshold, sensor
and general parameters.
Table 22
– Mnemonics for write commands
Read and Write Commands
Description
RSA
Read Serial Address
Read address for RS485.
SSA
Set Serial Address
Set address for RS485.
RGP
Read General Parameter
Read instrument settings.
SGP
Set General Parameter
Set instrument settings.
RGC
Read Gas Correction
Read type of gas correction factor for a measurement channel.
SGC
Set Gas Correction
Set type of gas correction factor for a measurement channel.
RSC
Read Sensor Control
Read type of sensor control for a measurement channel.
SSC
Set Sensor Control
Set type of sensor control for a measurement channel.
RSP
Read Setpoint
Read switching thresholds for a channel.
SSP
Set Setpoint
Set switching thresholds for a channel.
Table 23
– Mnemonics for read and write commands
CAUTION:
All changes to values are only permanently saved to the EEPROM after
having issued the write command SAC. For this reason after having set all
values, save the current configuration through the SAC write command before
restarting the instrument.