6/102
17200234_002_C1/2016 - © Leybold
602494EA.DOCX
1.4
Technical Data
General technical data
Type / Element
Value / Description
Power supply
Mains voltage [VAC]
90 … 250
with 50/60 Hz
Power consumption [VA]
50
Overvoltage category
II
(EN 61010-1)
Protection class
I
(EN 61010-1)
Main switch
On the rear of the CMOVE 1250
Controller
Controller types
Auto (PI) adjustable
1 … 99
(1 = slow, 99 = fast)
PID adjustable
Gain Kp: 0.1 … 100.0
Reset time Tn: 0.0 … 3600.0 s
Derivative time Tv: 0.0 … 3600.0 s
Control accuracy
1
0.5 % F.S.
Scan rate of pressure sensors [ms]
<100
Update rate [ms]
Inputs/outputs
<100
MOVE X
<100
MOVE 1250
<500
Display
LCD 64×128 Pixel
Measurement units (selectable)
Pressure
mbar, Torr, Pa, mV
Flow
mbar l/s, Torr l/s, Pa l/s, %, mV
1
Valid for sensor setting 0 … +10 V lin and output AO 2.