Device parameters and interfaces
84
AMS 200
Leuze electronic
8
.2.5
Changing the SSI default settings via the display
Notice
!
For basic operation of the display please refer to chapter 6.
In order to change the parameters please activate parameter enabling.
The SSI interface remains active even during parameter enabling. Changes to parameters
have an immediate effect.
8
.2.6
SSI parameter display
Figure 8.10: SSI parameter display
E
S
C
E
S
C
E
S
C
E
S
C
E
S
C
E
S
C
E
S
C
E
S
C
Activation of SSI interface ON/OFF
Default: ON
Data coding: Gray or binary
Default: Gray coding
Transmission mode:
24-bit, 25-bit or 26-bit measurement value
Default: 24-bit
Measurement value resolution:
0.001 mm, 0.01 mm, 0.1 mm, 1 mm or
10 mm
Default: 0.1 mm
Error bit: ON or OFF,
default: ON
Error bit is set in the event of overflow or intensity (ATT) or excess temperature (TMP) or laser prefailure
warning (LSR) or plausibility error (PLB) or hardware error (ERR).
Default: Plausibility error (PLB) or hardware error (ERR).
➀
➀
Update rate:
0.2 ms or 1.6 ms
(independent of the clock frequency!)
Default: 1.6 ms