5-1
Chapter 5 Alarms
Alarms are triggered by physiological parameters that appear
abnormal or by technical problems of the monitor. When an alarm
occurs, the monitor indicates it through visual and audible alarm
indications.
5.1 Safety Information
Warning
Before monitoring a new patient, always check that the monitor can work
properly, the alarm system works properly, and the alarm settings are
appropriate for the patient before starting the monitoring.
Warning
In order to ensure that the operator can accurately identify the alarms, it is
recommended that the distance between the operator and the monitor
should not exceed 4 meters. If the alarm event needs to be clearly
distinguished, it is recommended that the distance between the operator
and the monitor should not be more than 1 m (there should be no obstacle
within the visual effective distance above).
Warning
Do not set the alarm limits beyond the measurement ranges, which may
cause the alarm system to become ineffective.
Warning
For the same or similar equipment used in any separate area, there is a
potential hazard if different alarm presets are used.
Caution
When the alarm system is powered off, or loses all power, the monitor will
save the alarm settings and alarm logs if the power off time is not more
than 30s. The stored alarm information does not change with the power off
time.
Note
The alarm system function of the monitor can be checked and verified in
the demo mode.
Содержание AIView 12V
Страница 1: ...AIView Series Patient Monitor Operator s Manual...
Страница 2: ......
Страница 6: ...IV This page intentionally left blank...
Страница 24: ...1 7 This page intentionally left blank...
Страница 26: ...2 2...
Страница 31: ...2 7 2 6 2 Rear View 1 2 3 4 1 Handle 2 Device label 3 Equipotential grounding terminal 4 AC power connector...
Страница 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Страница 35: ...2 11 This page intentionally left blank...
Страница 44: ...3 9 1 2 3 4 5 8 9 7 6...
Страница 55: ...3 20 This page intentionally left blank...
Страница 71: ...5 11 This page intentionally left blank...
Страница 78: ...6 7 This page intentionally left blank...
Страница 106: ...8 5 This page intentionally left blank...
Страница 112: ...9 6 This page intentionally left blank...
Страница 119: ...10 7 This page intentionally left blank...
Страница 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Страница 165: ...15 12 Other sources of interference if any...
Страница 166: ...15 13 This page intentionally left blank...
Страница 181: ...17 8 Select Range to view the normal range of each parameter Select Unit to view the unit of each parameter...
Страница 185: ...17 12 Select Range to view the normal range of each parameter Select Unit to view the unit of each parameter...
Страница 194: ...17 21 This page intentionally left blank...
Страница 204: ...19 5 This page intentionally left blank...
Страница 210: ...20 6 This page intentionally left blank...
Страница 229: ...23 10 This page intentionally left blank...
Страница 248: ...25 8 This page intentionally left blank...
Страница 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Страница 283: ...D 2 This page intentionally left blank...
Страница 289: ...E 6 less than 3V m...