94
Check & Adjust
4.2
Preparation
Before starting to work, the instrument has to become acclimatised to the ambient
temperature. Approximately two minutes per °C of temperature difference from
storage to working environment but at least 15 min should be taken into account.
Note, that even after good adjustment of the ATR, the crosshairs might not be posi-
tioned exactly on the centre of the prism after an ATR measurement has been
executed. This is a normal effect. To speed up the ATR measurement, the telescope
is normally not positioned exactly on the centre of the prism. The small rest devia-
tions, the ATR offsets are measured individually for each measurement and
Before determining the instrument errors, the instrument has to be levelled-
up using the electronic level.
to access
STATUS Level & Laser
Plummet, Level
page.
The tribrach, the tripod and the underground should be very stable and
secure from vibrations or other disturbances.
The instrument should be protected from direct sunlight in order to avoid
thermal warming.
It is also recommended to avoid strong heat shimmer and air turbulence.
The best conditions are usually early in the morning or with overcast sky.
Содержание TPS1200+ Series
Страница 1: ...Leica TPS1200 User Manual Version 7 0 English...
Страница 11: ...Table of Contents TPS1200 11 8 International Limited Warranty Software License Agreement 202 Index 204...
Страница 35: ...User Interface TPS1200 35 SHIFT Pages up SHIFT Pages down Keys Description...
Страница 89: ...Operation TPS1200 89...
Страница 145: ...Safety Directions TPS1200 145 a Laser beam b Exit for laser beam b TPS12_048 a...
Страница 167: ...Technical Data TPS1200 167 Characteristics Principle Digital image processing Type Infrared laser...
Страница 182: ...182 TPS1200 Technical Data Antenna Type Patch antenna omnidirectional Gain 2 dBi Connector SMB...
Страница 213: ...Index TPS1200 213...