
209
Summarized manual for FCS or other point measurements
19
Summarized manual for FCS or other point
measurements
19.1
Prerequisites
The following description assumes that the following prerequisites are met:
• SymPhoTime is started and a workspace is loaded.
• LAS AF was started.
• The right objective has been chosen and the correction ring adjusted to the specimen
used (
and
• Your specimen is already placed onto the microscope and brought into focus. If you are
measuring in solution, it is assumed that the laser is focused into the solution (see
).
• The selected excitation lasers are switched on and activated (
and
).
• The correct emission filter(s) has/have been placed in your detector unit (
).
• For beam splitter systems: The correct beam splitter for your wavelength has been
selected in LAS AF.
• With AOBS systems: The acousto-optical beam splitter is in fluorescence mode.
19.1.1
Choosing the Location for the FCS Measurement
1. Start the FCS Wizard in LAS AF.
2. Scan an image. Also use the
Setup Imaging
step in the FCS Wizard.
3. Select the position at which you want to run a point measurement (
).
4. Start the FCS test mode in LAS AF by selecting the image acquisition button
Run FCS Test
under the
Setup FCS
step in the FCS Wizard (
Normally, the SymPhoTime software will start the data acquisition automatically.
However, you can also start the FCS data acquisition by selecting the
Acquire Point
Measurements
button in the SymPhoTime software.
The
Measurement Preview
dialog in SymPhoTime opens on the
FCS
tab (
5. The display of the cross-trace and the two autocorrelation traces can be selected using
the check boxes above the trace.
6. The update time can be set in the
Refresh
field. Together with the average count rate,
the G(0) value and the molecular brightness (
Counts/Molec.
) are shown for the individual
channels.
The molecular brightness is calculated by G(0) times the average count rate and displays
the mean fluorescence count rate per dye molecule. The system is optimized (for example
by adjusting the laser intensity and the objective collar) when maximum molecular
brightness is attained.
In the
Time Trace
, right) the changes of the count rate over time can be
observed. The binning (
resolution
) should be set to 1 ms.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......