F(L)CS Data Acquisition
• The user does not have access to the
Total number of measurements
field. It serves to
visualize the number of FCS measurements carried out (
). This number
is the product of
Number of defined points
times
Repetition at point
times
Number of
cycles
.
The maximum number of FCS measurements in a measurement series is 1000.
Start the measurement series in LAS AF by clicking the
Run FCS
image acquisition button.
18.4.4
Definition of the File Names Transferred to SymPhoTime
during the
FCS Measurement
You can specify the name of the FCS data file used in SymPhoTime in the
Setup
tab in the
Definition of point measurement
field (
).
The following functions are available:
• Whenever a series of FCS measurements is started, SymPhoTime generates a folder that
houses all FCS measurement files of this series. The name of the folder is composed as
follows:
•
BaseName_Run
Base name
: Enables you to specify the base folder and file name used in SymPhoTime
). If the user does not enter a name,
BaseName
will be
automatically used.
Run
: This is a suffix that is added to the file names transferred in SymPhoTime.
In SymPhoTime, no two folders may have the same name. Therefore, the number
given in the
Run
suffix is automatically increased by one whenever a measurement
series is started. The
Run
suffix can be changed by the user (
).
• The file name uses the same
BaseName
and
Run
components as the folder. Additionally,
LAS AF will automatically add some more suffixes to the file name that contain the
following information:
•
Cycle
•
Measuring point
•
Repetitions at this point
The file name structure is as follows:
•
BaseName_Run_Cycle_Measuring point_Repetition
•
The
Measuring point
number is identical to the number used in the image display.
These additional suffixes cannot be modified by the user.
• If
Image after each __ cycles
is enabled, an image or stack is automatically saved in
LAS AF with the following name:
BaseName_Run_Cycle
Do not use the same name
Do not enter the same
BaseName_Run
name twice. This will create an
error message.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......