F(L)CS Data Acquisition
Figure 177:
Measurement Preview in SymPhoTime: Allocation of the FCS channels
Defaults can be saved using the
Set Defaults
button.
In addition to the count rate (cps), the count rate per molecule (cpm) is displayed in the
count rate monitor in LAS AF. It is calculated from the online correlation. Since FCS is based
on fluctuation analysis, it is normal for these values to fluctuate. The higher this value, the
better the signal/noise ratio of the correlation curve. That means this value should be
maximized (by selecting an appropriate fluorescent dye, a suitable laser intensity, a suitable
dye concentration, the correct setting for the objective's correction ring, etc.).
To prevent saturation effects, select the laser intensity with which you can reach
approximately two-thirds of the maximum count rate per molecule. If you then observe
photobleaching or an excessively high triplet fraction, continue to reduce the laser intensity.
18.3.7
Loading and Saving FCS-specific Instrument Parameter Settings
FCS-specific instrument parameter settings (IPS) are organized in a different way than
image acquisition settings. Therefore, they are managed separately from the image
acquisition settings. On the one hand, this means that FCS instrument parameter settings
can be loaded and saved only in the
Setup FCS
step; they are not available anywhere else
(
). On the other hand, it means that image acquisition settings are not available in
the
Setup FCS
step. They must be loaded and saved in the
Setup Imaging
step or outside of
the FCS Wizard.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......