174
Summarized manual for FLIM experiments
17.12.2
Running IRF Measurement
Carry out the following steps to measure the IRF:
1. In the FLIM Wizard in LAS AF, go to step
Setup FLIM
.
2. Start the scanning operation in test mode by selecting the image acquisition button
Run
FLIM Test
. Normally, the SymPhoTime software will start the data acquisition
automatically. However, you can also start the FLIM data acquisition by selecting the
Acquire LSM Measurements
button.
3. Enable the
Time Trace
tab in the measurement preview window.
4. Adjust the focus of the microscope in order to maximize the displayed count rate.
5. Set the count rate to approximately 50,000 count/s.
6. The laser output power must not be changed on the laser driver itself, but only on the
laser coupling module (LCU) or on the scaffold on the laser head.
Changing the laser driver settings typically changes the shape and temporal position of the
laser pulse. You can use the driver settings to optimize the pulse shape of the laser for
shortest IRF width.
7. You can now test the fading behavior of your specimen in the oscilloscope window in the
SymPhoTime software. For this purpose, quit the preview and select
Show
Oscilloscope...
.
8. Select the
TCSPC Histogram
tab, select
Start
and monitor the fading of the fluorescence
as shown in
. In the oscilloscope window, the time width per channel can also
be set.
Data is not saved.
Data from a scan in test mode is not saved.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......